Journal of cataract and refractive surgery · 2009
Effect of hydrodynamic parameters on corneal endothelial cell loss after phacoemulsification.
СтатьяРандомизированное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Baradaran-Rafii Alireza, Rahmati-Kamel Mohsen, Eslani Medi, Kiavash Victoria, Karimian Farid
Аннотация
PURPOSE: To evaluate the effect of power, vacuum, and flow rate on endothelial cell loss after phacoemulsification.
SETTING: Labbafinejad Medical Center Medical Center, Tehran, Iran.
METHODS: In a prospective randomized clinical trial, phacoemulsification was performed in 2 groups (high vacuum and low vacuum) with 3+ nuclear sclerosis. The stop-and-chop technique was used with the Sovereign machine. Machine parameters during the chop stage were vacuum 400 mm Hg in the high-vacuum group and 200 mm Hg in the low-vacuum group and flow rate, 40 cc/min and 20 cc/min, respectively. Endothelial cell density preoperatively before and 1, 6, and 12 weeks postoperatively were compared.
RESULTS: Each group comprised 30 eyes. The mean US power was 9.2% +/- 4.3% (SD) in the low-vacuum group and 13.1% +/- 4.6% in the high-vacuum group (P = .001) and the mean phaco time, 1.28 +/- 1.0 minutes and 0.88 +/- 0.6 minutes, respectively (P = .04). Total US energy and total fluid consumed were similar between groups. After 12 weeks, the mean endothelial cell loss was 9.0% +/- 4.0% in the low-vacuum group and 9.6% +/- 4.6% in the high-vacuum group (P = .6). There was a relationship between total US energy and endothelial loss (P<.001); however, total fluid volume was not a significant predictor (P = .19).
CONCLUSIONS: Vacuum level did not have a significant effect on total US energy or total fluid consumed during phacoemulsification. There was a strong relationship between total US energy and endothelial cell loss but not between total infused fluid and endothelial cell loss.
SETTING: Labbafinejad Medical Center Medical Center, Tehran, Iran.
METHODS: In a prospective randomized clinical trial, phacoemulsification was performed in 2 groups (high vacuum and low vacuum) with 3+ nuclear sclerosis. The stop-and-chop technique was used with the Sovereign machine. Machine parameters during the chop stage were vacuum 400 mm Hg in the high-vacuum group and 200 mm Hg in the low-vacuum group and flow rate, 40 cc/min and 20 cc/min, respectively. Endothelial cell density preoperatively before and 1, 6, and 12 weeks postoperatively were compared.
RESULTS: Each group comprised 30 eyes. The mean US power was 9.2% +/- 4.3% (SD) in the low-vacuum group and 13.1% +/- 4.6% in the high-vacuum group (P = .001) and the mean phaco time, 1.28 +/- 1.0 minutes and 0.88 +/- 0.6 minutes, respectively (P = .04). Total US energy and total fluid consumed were similar between groups. After 12 weeks, the mean endothelial cell loss was 9.0% +/- 4.0% in the low-vacuum group and 9.6% +/- 4.6% in the high-vacuum group (P = .6). There was a relationship between total US energy and endothelial loss (P<.001); however, total fluid volume was not a significant predictor (P = .19).
CONCLUSIONS: Vacuum level did not have a significant effect on total US energy or total fluid consumed during phacoemulsification. There was a strong relationship between total US energy and endothelial cell loss but not between total infused fluid and endothelial cell loss.