Confirmation of successful supraglottic airway device placement in neonates using a respiratory function monitor.

Dvorsky Robyn, Werther Tobias, Bibl Katharina, Schneider Michael, Binder Christoph, Habrina Lisa, Klebermaß-Schrehof Katrin, Kranebitter Veronika, Schmölzer Georg M, Berger Angelika, Wagner Michael

СтатьяПеревод в очереди

Аннотация

BACKGROUND

This study investigated the use of a respiratory function monitor (RFM) to guide the placement of a supraglottic airway device (SAD) in neonates during intensive care interventions. We hypothesized that using a RFM would decrease the number of attempts needed for a successful placement.

METHODS

This single-center pilot study was carried out at a tertiary NICU at the Medical University of Vienna. Patients were ventilated using a SAD during neurosurgical or endoscopic interventions. A RFM was either hidden (but recording) or visible to providers during SAD placement. Feedback from the RFM was used to assess correct/incorrect placement and optimize ventilation quality. The parameter leakage was used for assessment: if leak was <30%, correct placement was assumed. The primary outcome was the number of attempts until correct placement. Secondary outcomes included ventilation parameters recorded by the RFM and the duration of SAD placement.

RESULTS

Six patients were included in this pilot trial. Using a RFM to guide SAD placement led to fewer attempts (median attempts: 3 [hidden] vs. 1 [visible]). Furthermore, using the RFM, necessary adaptations were made to the SAD position to decrease leakage (mean leakage: 74.8% [hidden] vs. 17.8% [visible]), subsequently endoscopy after insertion of SAD using the RFM then confirmed anatomically correct position.

CONCLUSION

This pilot study indicated that a RFM might be useful to provide guidance during SAD placement.

IMPACT STATEMENT: Feedback from a RFM reliably indicated correct anatomical placement of a SAD by correlating low leakage values with proper SAD positioning. RFM guidance could improve neonatal airway management, reducing procedural time and number of attempts. We present promising preliminary results. Further research is needed to confirm these findings.

Для цитирования: Dvorsky Robyn, Werther Tobias, Bibl Katharina, Schneider Michael, Binder Christoph, Habrina Lisa, Klebermaß-Schrehof Katrin, Kranebitter Veronika, Schmölzer Georg M, Berger Angelika, Wagner Michael. Confirmation of successful supraglottic airway device placement in neonates using a respiratory function monitor.. Pediatric research. 2025. DOI: 10.1038/s41390-025-03810-x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.