MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.clinbiomech.2011.05.012

Медицинская библиотека · Clinical biomechanics (Bristol, Avon) · 2011

Arthroscopic rotator cuff repair: a biomechanical comparison of the suture-bridge technique vs. a new transosseous technique using SutureButtons(®).

Tauber Mark, Hoffelner Thomas, Penzkofer Rainer, Koller Heiko, Zenner Juliane, Hitzl Wolfgang, Moroder Philipp, Resch Herbert

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

The suture-bridge technique using anchors as established transosseous-equivalent technique in arthroscopic rotator cuff repair was compared to a modified transosseous technique suitable for arthroscopic cuff repair.

METHODS

In 10 fresh-frozen matched pairs of human cadaveric shoulders (mean age 67.1, SD 8.5 years), two different surgical techniques of cuff repair were tested: Group 1, using the suture-bridge technique with suture anchors, and Group 2, using two transosseous tunnels with SutureButtons(®). Lateral row fixation was performed in both groups using knotless implants. Cyclic displacement to gap formation of 2 and 5mm, linear stiffness, yield load, ultimate load, and mode of failure were recorded.

FINDINGS: Gap formation at the tendon-to-bone interface of 2mm occurred after a mean of 219.5 (SD 590.7) cycles in Group 1 and after 750.0 (SD 1566.1) cycles in Group 2. Gap formation of 5mm occurred after 2331.6 (SD 2033.9) cycles (Group 1) and 2364.5 (SD 1994.2) cycles (Group 2), respectively. The yield and ultimate loads were 316.9 (SD 114.1)N and 375.9 (SD 131.2)N in Group 1, and 311.0 (SD 97.2)N and 363.8 (SD 107.6)N in Group 2, respectively. The linear stiffness was 40.3 (SD 10.4)N/mm in Group 1, and 41.6 (SD 13.2)N/mm in Group 2. There were no statistically significant intergroup differences.

INTERPRETATION: The new transosseous technique using SutureButtons(®) achieves equivalent biomechanical properties to the established suture-bridge technique using anchors. A tendentially reduced primary gap formation may be of importance for repair healing during the early phase of rehabilitation.

Для цитирования: Tauber Mark, Hoffelner Thomas, Penzkofer Rainer, Koller Heiko, Zenner Juliane, Hitzl Wolfgang, Moroder Philipp, Resch Herbert. Arthroscopic rotator cuff repair: a biomechanical comparison of the suture-bridge technique vs. a new transosseous technique using SutureButtons(®).. Clinical biomechanics (Bristol, Avon). 2011. DOI: 10.1016/j.clinbiomech.2011.05.012.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.