DOI 10.1016/j.pedneo.2022.05.012

Медицинская библиотека · Pediatrics and neonatology · 2022

Accuracy of transcutaneous bilirubin level measured by a JM-105 bilirubinometer.

Ohishi Akira, Kondo Miyuki, Fujita Tomoka, Baba Toru, Iijima Shigeo

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Transcutaneous bilirubin (TcB) measurement is useful, but dissociation with total serum bilirubin (TSB) is a clinical problem in measurement. We verified the accuracy of the latest version of the JM-105 jaundice meter.

METHODS

The TcB, TSB, and hematocrit (Hct) measurements obtained in the first 4 days of life in 2788 term neonates were analyzed.

RESULTS

When divided into 2-mg/dL classes, the difference between the TcB and TSB measurements did not change as TcB increased, but both overestimation and underestimation of TcB increased as TcB increased. At TcB greater than 11 mg/dL, inaccurate measurements with dissociation greater than 2 mg/dL exceeded 10% of the TcB measurements. The Hct value was associated with overestimation and underestimation.

CONCLUSION

To evaluate neonatal jaundice accurately, it is desirable to measure TSB by blood sampling before discharge from obstetrics or in the case of worsening jaundice on day 4 or 5 of life.

Для цитирования: Ohishi Akira, Kondo Miyuki, Fujita Tomoka, Baba Toru, Iijima Shigeo. Accuracy of transcutaneous bilirubin level measured by a JM-105 bilirubinometer.. Pediatrics and neonatology. 2022. DOI: 10.1016/j.pedneo.2022.05.012.