Total bilirubin assay differences may cause inconsistent treatment decisions in neonatal hyperbilirubinaemia.

Thomas David H, Warner Janet V, Jones Graham R D, Chung Jason Z Y, Macey David J, Screnci Antonella, Ryan Joshua B

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

To assess interlaboratory variability of total serum bilirubin (TSB) results in newborns. Initiated following a clinical incident in which a neonate was transferred to a tertiary hospital for treatment of severe hyperbilirubinemia but on arrival was reclassified into a lower risk category due to a 20% difference in TSB between laboratories.

METHODS

Fresh residual plasma samples from hospital-born infants were pooled to obtain 11 samples across a range of total bilirubin concentrations. Aliquots were light-protected and measured on 7 commercial platforms at 4 accredited medical laboratories. Data from The Royal College of Pathologists of Australasia Quality Assurance Programs' (RCPAQAP) Neonatal Bilirubin program was analysed.

RESULTS

Twenty-four to 30% difference in results for individual samples, largely due to calibration differences between assays. When interpreted according to guidelines, results from different platforms would have led to different clinical interventions in some cases. RCPAQAP results showed significant within-method bias but were not shown to be commutable with patient samples.

CONCLUSIONS

There are clinically significant method-dependent differences in TSB results from neonatal samples, consistent with our clinical incident. The differences are largely due to lack of standardisation of calibrator values. This has implications for healthcare resource use and possibly for the neurodevelopment of infants. Intervention is needed at a number of levels, including clinical reporting of incidents arising from discordant results, commitment by manufacturers to ensure metrological traceability of methods with sufficiently low uncertainty in the final measurements, and availability of commutable quality assurance material to monitor assay performance, especially at the clinical decision points for neonatal jaundice.

Для цитирования: Thomas David H, Warner Janet V, Jones Graham R D, Chung Jason Z Y, Macey David J, Screnci Antonella, Ryan Joshua B. Total bilirubin assay differences may cause inconsistent treatment decisions in neonatal hyperbilirubinaemia.. Clinical chemistry and laboratory medicine. 2022. DOI: 10.1515/cclm-2022-0749.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.