Spuriously elevated estradiol due to immunoassay interference in the workup of a suspected ovarian tumor: Diagnostic pitfalls and laboratory workflow recommendations.

Lupoli Gelsy Arianna, Polito Carmela, Jannuzzi Giuseppe, Fiorenza Mariano, Marino Ada, Di Meglio Lavinia, Salvatore Domenico, Terracciano Daniela

Клинический случайСтатьяПеревод в очереди

Аннотация

BACKGROUND

Falsely elevated estradiol levels due to analytical interference are rare but clinically relevant diagnostic pitfalls, particularly when results suggest an estrogen-producing tumor.

CASE PRESENTATION: We report the case of a 51-year-old woman with markedly increased estradiol concentrations inconsistent with her clinical presentation and hormonal status. Persistently elevated estradiol values, together with imaging findings and concern for granulosa cell tumor, contributed to the diagnostic workup of a suspected estrogen-producing ovarian tumor. Bilateral salpingo-oophorectomy was performed within a multifactorial preoperative assessment; histopathology showed no evidence of malignancy.

LABORATORY INVESTIGATION: Postoperatively, estradiol remained markedly elevated despite gonadotropin values consistent with surgical menopause. Re-measurement on an alternative immunoassay platform revealed low postmenopausal estradiol levels, whereas the original competitive chemiluminescent assay continued to show persistently elevated values. Polyethylene glycol precipitation produced a substantial reduction in estradiol concentration, supporting immunoglobulin-mediated analytical interference, most likely due to heterophile antibodies.

DISCUSSION: This case highlights the vulnerability of competitive immunoassays to analytical interference, which may paradoxically generate falsely elevated results through signal suppression. Clinical-biochemical discordance should prompt timely communication between clinicians and laboratory specialists, repeat testing, alternative-platform measurement, and targeted interference studies.

CONCLUSION

Recognition of analytical interference is essential to prevent misdiagnosis and avoid unnecessary or potentially harmful interventions. We propose a practical diagnostic workflow to support the identification and management of discordant estradiol results in routine laboratory practice.

Для цитирования: Lupoli Gelsy Arianna, Polito Carmela, Jannuzzi Giuseppe, Fiorenza Mariano, Marino Ada, Di Meglio Lavinia, Salvatore Domenico, Terracciano Daniela. Spuriously elevated estradiol due to immunoassay interference in the workup of a suspected ovarian tumor: Diagnostic pitfalls and laboratory workflow recommendations.. Clinical biochemistry. 2026. DOI: 10.1016/j.clinbiochem.2026.111144.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.