Novel open-source calculator for lens tilt and decentration using SS-AS-OCT: assessment of interrater agreement and reproducibility.

Seiller-Tarbuk Konstantin, Puchner Stefan, Regitnig Paul, Kahraman Guenal, Amon Michael

СтатьяПеревод в очереди

Аннотация

PURPOSE: To introduce an open-source calculator for quantifying lens tilt and decentration using swept-source anterior segment optical coherence tomography (SS-AS-OCT) images, and to evaluate its reproducibility and independence from evaluator experience.

SETTING

Department of Ophthalmology, Hospital of the Brothers of St. John of God, Vienna, Austria.

DESIGN

Prospective interrater agreement study.

METHODS

Patients scheduled for cataract surgery were screened for eligibility. Inclusion required age 21 years or older, written informed consent, and adequate fixation. Exclusion criteria included physical immobility or other limitations preventing imaging, prior pseudophakia in the fellow eye, contraindications to pharmacologic mydriasis, or tremor. Imaging was performed under pharmacologically induced mydriasis using SS-AS-OCT. Lens tilt and decentration were independently assessed by a novice (medical student) and an expert (ophthalmologist) using a self-developed open-source calculator.

RESULTS

175 eyes from 88 patients, including 30 men (34%) and 58 women (66%) with a mean age of 72 years (range 30 to 91), were included. According to expert measurements, lens tilt averaged 5.65 degrees, with a consistent inferotemporal orientation (5.06 degrees temporal, 1.81 degrees inferior). Decentration averaged 146 µm and was primarily directed temporally (79.51 µm), with only a minor inferior component (2.93 µm). The intraclass correlation coefficient (ICC) was 0.972 (95% CI 0.962-0.979) for tilt and 0.914 (95% CI 0.886-0.935) for decentration, reflecting excellent agreement between raters.

CONCLUSIONS

The evaluated open-source calculator enables accurate and reproducible measurement of lens tilt and decentration using SS-AS-OCT images. In this study, interrater agreement was excellent for both parameters, with ICC values above 0.9 and no evidence of systematic measurement bias.

Для цитирования: Seiller-Tarbuk Konstantin, Puchner Stefan, Regitnig Paul, Kahraman Guenal, Amon Michael. Novel open-source calculator for lens tilt and decentration using SS-AS-OCT: assessment of interrater agreement and reproducibility.. Journal of cataract and refractive surgery. 2026. DOI: 10.1097/j.jcrs.0000000000001841.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.