MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s00402-023-05024-4

Медицинская библиотека · Archives of orthopaedic and trauma surgery · 2023

Digital measurement of anterolateral knee laxity using strain sensors.

Mayr Hermann O, Rosenstiel Nikolaus, Prakash Karthika S, Comella Laura M, Woias Peter, Schmal Hagen, Seidenstuecker Michael

СтатьяПеревод в очереди

Аннотация

PURPOSE: The ambition of the research group was to develop a sensor-based system that allowed the transfer of results with strain sensors applied to the knee joint. This system was to be validated in comparison to the current static mechanical measurement system. For this purpose, the internal rotation laxity of the knee joint was measured, as it is relevant for anterolateral knee laxity and anterior cruciate ligament (ACL) injury.

METHODS

This is a noninvasive measurement method using strain sensors which are applied to the skin in the course of the anterolateral ligament. The subjects were placed in supine position. First the left and then the right leg were clinically examined sequentially and documented by means of an examination form. 11 subjects aged 21 to 45 years, 5 women and 6 men were examined. Internal rotation of the lower leg was performed with a torque of 2 Nm at a knee flexion angle of 30°.

RESULTS

Comparison of correlation between length change and internal knee rotation angle showed a strong positive correlation (r = 1, p < 0.01). Whereas females showed a significant higher laxity vs. males (p = 0.003).

CONCLUSIONS

The present study showed that the capacitive strain sensors can be used for reproducible measurement of anterolateral knee laxity. In contrast to the previous static systems, a dynamic measurement will be possible by this method in the future.

Для цитирования: Mayr Hermann O, Rosenstiel Nikolaus, Prakash Karthika S, Comella Laura M, Woias Peter, Schmal Hagen, Seidenstuecker Michael. Digital measurement of anterolateral knee laxity using strain sensors.. Archives of orthopaedic and trauma surgery. 2023. DOI: 10.1007/s00402-023-05024-4.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.