Comparison of Cup Setting Angle Accuracy between Computed Tomography-Based and Computed Tomography-Free Navigation in the Same Patients with Crowe's Classification I or II Hip Dysplasia.

Kaku Nobuhiro, Tagomori Hiroaki, Tsumura Hiroshi

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUD: In total hip arthroplasty, the cup setting angle may affect the postoperative results. In recent years, both computed tomography-based navigation and computed tomography-free (imageless) navigation have been reported to produce high accuracy in cup installation; however, no direct comparison between these two methods has been performed. The present study aimed to directly compare the cup installation angle accuracy between computed tomography-based navigation and computed tomography-free navigation in patients with Crowe's classification stage I or II dysplastic osteoarthritis and to examine the factors affecting the cup installation accuracy.

METHODS

Using both navigation systems for the same technique, primary total hip arthroplasty was performed by the same surgeon in 36 patients. A cup was installed using computed tomography-based navigation, and the installed cup was measured again using computed tomography-free navigation. We compared the error between the target angle and the intraoperative installation angle for each navigation method by performing statistical analyses.

RESULTS

For computed tomography-based navigation, errors in the inclination and the anteversion angles compared to the target angle were 3.14° ± 1.55° and 1.47° ± 0.99°, respectively. For computed tomography-free navigation, the inclination and anteversion angle errors were significantly larger, i.e., 6.84° ± 4.78° and 5.43° ± 5.22°, respectively (p < 0.01). The inclination and anteversion angles of computed tomography-free navigation were correlated, and there were no significant factors influencing the error.

CONCLUSIONS

Computed tomography-based navigation is more accurate for cup installation than computed tomography-free navigation. When using computed tomography-free navigation, it is necessary to add technical schemes before and during surgery to improve the cup installation accuracy.

Для цитирования: Kaku Nobuhiro, Tagomori Hiroaki, Tsumura Hiroshi. Comparison of Cup Setting Angle Accuracy between Computed Tomography-Based and Computed Tomography-Free Navigation in the Same Patients with Crowe's Classification I or II Hip Dysplasia.. Clinics in orthopedic surgery. 2021. DOI: 10.4055/cios20145.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.