MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1053/j.ajkd.2019.05.023

Медицинская библиотека · American journal of kidney diseases : the official journal of the National Kidney Foundation · 2019

An Implanted Blood Vessel Support Device for Arteriovenous Fistulas: A Randomized Controlled Trial.

Karydis Nikolaos, Bevis Paul, Beckitt Timothy, Silverberg Daniel, Halak Moshe, Calder Francis

СтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

RATIONALE &

OBJECTIVE

Reducing turbulent blood flow through dialysis arteriovenous fistulas (AVFs) and radial stretching of their venous wall may attenuate hyperplasia and stenosis and improve AVF outcomes in hemodialysis patients. The goal of this study was to evaluate the safety and efficacy of the VasQ implant, which intervenes on these mechanisms by physically supporting the surgical arteriovenous anastomosis.

STUDY

DESIGN

Prospective, randomized, controlled, multicenter study.

SETTINGS &

PARTICIPANTS

40 consecutive patients with kidney failure referred for creation of a brachiocephalic fistula in 4 vascular access centers in the United Kingdom and Israel.

INTERVENTIONS: AVF surgical creation with placement of the VasQ implant (treatment) versus AVF placement without the implant (control).

OUTCOMES: Safety assessed as percentage of severe device-related adverse events was the primary outcome. Secondary outcomes were efficacy assessments including: (1) AVF maturation at 3 months, defined as cephalic vein diameter≥5mm and flow≥500mL/min; (2) functional cumulative patency, defined as successful 2-needle cannulation for two-thirds or more of all dialysis runs for 1 month in study participants receiving dialysis; (3) cephalic vein diameter and blood flow; and (4) primary and cumulative patency at 6 months.

RESULTS

No severe device-related adverse events were observed. There was no significant difference in maturation at 3 months or primary patency at 6 months between treatment and control (85% vs 80% and 80% vs 66%). Significantly larger vein luminal diameters were observed in the treatment group versus controls at 3 and 6 months (8.27±2.2 vs 6.69±1.8mm [P=0.03] and 9.6±2.5 vs 7.56±2.7mm [P=0.03]). Functional patency at 6 months was significantly greater in the treatment group (100% vs 56% [P = 0.01]).

LIMITATIONS: Small sample size, limited power for secondary end points.

CONCLUSIONS

No safety signals were detected for the VasQ external support of brachiocephalic AVFs. Higher functional patency and vein luminal diameters were achieved with the device at 3 and 6 months. VasQ may safely intervene on mechanisms associated with the disturbed hemodynamic profile in the juxta-anastomotic region.

FUNDING: Funded by Laminate Medical Technologies Ltd.

TRIAL REGISTRATION: Registered at ClinicalTrials.gov with study number NCT02112669.

Для цитирования: Karydis Nikolaos, Bevis Paul, Beckitt Timothy, Silverberg Daniel, Halak Moshe, Calder Francis. An Implanted Blood Vessel Support Device for Arteriovenous Fistulas: A Randomized Controlled Trial.. American journal of kidney diseases : the official journal of the National Kidney Foundation. 2019. DOI: 10.1053/j.ajkd.2019.05.023.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.