Concordance between IDEXX Legiolert® (liquid culture assay) and plate culture (ISO 11731:2017) for the detection and quantification of Legionella pneumophila in water samples.

Karia K, Yui S, Muzslay M, Ali S

СтатьяСравнительное исследованиеОценочное исследованиеПеревод в очереди

Аннотация

BACKGROUND

Legionella pneumophila is a water-borne bacterium that can cause Legionnaires' disease. Legiolert® (IDEXX, USA) is a low-labour liquid culture assay for the detection and enumeration of L. pneumophila (SG1-15) from water.

AIM: To analyse concordance between Legiolert and ISO 11731:2017 plate culture method (membrane filtration and culture on selective agars) using hospital water samples (N = 100).

METHODS

Incubation was at 39 °C and 36 °C, respectively, for seven days, followed by most-probable enumeration for Legiolert and subculturing and serogrouping of suspected Legionella colonies, with plate culture.

FINDINGS: L. pneumophila (SG1-15) was isolated from 25 out of 100 samples when using Legiolert or plate culture. Fourteen additional Legiolert samples tested positive for L. pneumophila; analysis of the same samples by plate culture was negative (12 out of 14) or yielded only Legionella rubrilucens (two out of 14; confirmed via matrix-assisted ionization/desorption time-of-flight mass spectrometry). L. pneumophila was not captured from Quanti-Tray/Legiolert pouch wells of these positive samples after subculture of puncture aliquots on buffered charcoal yeast-extract agar. Both methods in concordance did not detect L. pneumophila in 61 out of 100 samples.

CONCLUSION

Legiolert and plate culture are both satisfactory methods to detect L. pneumophila from water samples, and both to detect isolated L. pneumophila in 25% of the sample population. Legiolert provides a faster time to result, and is less resource-demanding and labour-intensive; however, there may be a low risk of cross-reactivity with other organisms. Both methods are suitable for the analysis of water in healthcare settings, where the monitoring of L. pneumophila is imperative in preventing cases of Legionnaires' disease.

Для цитирования: Karia K, Yui S, Muzslay M, Ali S. Concordance between IDEXX Legiolert® (liquid culture assay) and plate culture (ISO 11731:2017) for the detection and quantification of Legionella pneumophila in water samples.. The Journal of hospital infection. 2023. DOI: 10.1016/j.jhin.2023.10.019.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.