DOI 10.1038/s41433-019-0509-9

Медицинская библиотека · Eye (London, England) · 2019

The infrared reflex: a potential new method for congenital cataract screening.

Duret Amedine, Humphries Robert, Ramanujam Sushmita, Te Water Naudé Anna, Reid Conor, Allen Louise E

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

To compare the accuracy of infrared (IR)-reflex assessment using a prototype imaging device to standard non-mydriatic red-reflex screening with direct ophthalmoscope (DO) in the diagnosis of neonatal and childhood cataract.

METHODS

The comparison of the techniques was made in two distinct cohorts: in the first, newborns underwent IR and red-reflex testing by a medical student, with results compared to a reference red-reflex examination by an experienced midwife. In the second, an enriched cohort of children attending a specialist paediatric ophthalmology clinic had IR and red-reflex testing by a medical student to reference examination by a paediatric ophthalmologist. The medical students were considered inexperienced screeners due to their limited exposure to ophthalmology. The sensitivity and specificity of the IR and red-reflex assessments in respect to reference examination were calculated. Diagnostic accuracy was compared in Caucasian and non-Caucasian eyes.

RESULTS

IR and red-reflex imaging were possible in all 180 neonatal eyes examined. A total of 5% of newborn eyes were found to have embryological remnants in the anterior segment of the eye with IR-reflex imaging which were not detected on reference red-reflex examination. IR-reflex assessment had significantly better sensitivity (100 vs 71%, p < 0.05) and specificity (100 vs 63%, p < 0.01) than red-reflex assessment in the diagnosis of childhood cataract. Red-reflex specificity was particularly poor in non-Caucasian eyes compared to Caucasian eyes (32 vs 72%, p < 0.05).

CONCLUSION

This pilot study indicates that IR-reflex imaging has the potential to improve the diagnostic accuracy of eye screening for cataract by inexperienced healthcare staff, particularly in non-Caucasian children.

Для цитирования: Duret Amedine, Humphries Robert, Ramanujam Sushmita, Te Water Naudé Anna, Reid Conor, Allen Louise E. The infrared reflex: a potential new method for congenital cataract screening.. Eye (London, England). 2019. DOI: 10.1038/s41433-019-0509-9.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.