DOI 10.4103/apc.apc_159_17

Медицинская библиотека · Annals of pediatric cardiology · 2018

Validation of the accuracy of handheld echocardiography for diagnosis of congenital heart disease.

Ali Sulafa, Bushari Tajudeen

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Handheld echocardiography (HHE) has been increasingly used for rheumatic heart disease screening and in adult emergency room settings.

AIMS: This study aimed to validate the accuracy of HHE in the diagnosis of congenital heart disease (CHD).

SETTINGS AND DESIGN METHODOLOGY: This is a prospective study carried out at three pediatric cardiology outpatient clinics in Khartoum, Sudan.All patients with suspected CHD were evaluated by clinical examination followed by HHE, performed using a modified segmental approach. Then, a complete study was performed using a standard echocardiography machine. The results were then compared using appropriate statistical tools.

RESULTS

A total of eighty cases were included with the following diagnoses either in isolation or combination: ventricular septal defect (n = 23), atrial septal defect (n = 10), pulmonary stenosis (n = 7), tetralogy of Fallot (n = 7), patent ductus arteriosus (n = 6), atrioventricular septal defect (n = 6), transposition of the great arteries (n = 6), and other diagnoses (n = 15). Agreement between HHE and SE was excellent both for visualizing heart segments (κ =77%-100% with a mean of 92.9%) and for diagnosis of CHD (κ =66%-100% with a mean of 91.7%). The sensitivity of HHE was 69.2%-100% (mean = 90.2%) and specificity was 98.5%-100% (mean = 99.3%).

CONCLUSION

This study supports extending the utility of HHE in children for screening of CHD in addition to its current role in rheumatic heart disease screening.

Для цитирования: Ali Sulafa, Bushari Tajudeen. Validation of the accuracy of handheld echocardiography for diagnosis of congenital heart disease.. Annals of pediatric cardiology. 2018. DOI: 10.4103/apc.apc_159_17.