DOI 10.3390/diagnostics14242870
Local Myoelectric Sensing During Human Colonic Tissue Perfusion.
Аннотация
Objectives
Anastomotic leakage (AL) is one of the most devastating complications after colorectal surgery. The verification of the adequate perfusion of the anastomosis is essential to ensuring anastomosis integrity following colonic resections. This study aimed to evaluate the efficacy of measuring the electrical activity of the colonic muscularis externa at an anastomosis site for perfusion analysis following colorectal surgery.
Methods
Strips of human isolated colon were maintained in a horizontal tissue bath to record spontaneous contractions and myoelectric activity and spike potentials (using a bipolar electrode array for the wireless transmission of myoelectric data-the xBar system) from the circular muscle. Intraoperative myoelectric signal assessment was performed by placing the electrode array on the colon prior to and following mesenteric artery ligation, just prior to colonic resection.
Results
In human isolated colon, the amplitude, duration, and frequency of contractions were inhibited during hypoxia by >80% for each measurement, compared to control values and time-matched oxygenated muscle. Intraoperative (N = 5; mean age, 64.8 years; range, 54-74 years; 60% females) myoelectric signal assessment revealed a decline in spike rate following arterial ligation, with a mean reduction of 112.64 to 51.13 spikes/min (p < 0.0008). No adverse events were observed during the study, and the device did not substantially alter the surgical procedure.
Conclusions
The electrical and contraction force of the human colon was reduced by ischemia, both in vitro and in vivo. These preliminary findings also suggest the potential of the xBar system to measure such changes during intraoperative and possibly postoperative periods to predict the risk of anastomotic viability as a surrogate of evolving dehiscence.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.