Validation of MICA Legionella for Enumeration of Legionella pneumophila in Sanitary Waters and Cooling Tower Waters: AOAC Performance Tested MethodSM 032201.

Passot Fanny, Peslier Sabine, Benzinger M Joseph, Blackburn Jonathan, Thompson Wesley, Bastin Benjamin, Dumont Audrey, Dukan Sam

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Frequent testing for Legionella concentration in water is required by most health risk monitoring organizations worldwide. Domestic hot water and cooling tower water networks must be regularly controlled to prevent Legionnaires' disease, a potentially deadly lung infection. MICA Legionella is the fastest culture-based detection method for all serogroups of Legionella pneumophila, with automatic enumeration in 48 h and no need for confirmation.

OBJECTIVE

This study compares the performance and robustness of MICA Legionella with the reference method ISO 11731:2017 for the enumeration of culturable L. pneumophila.

METHODS

MICA Legionella and ISO 11731:2017 results were compared for domestic hot water and cooling tower water. Inclusivity and exclusivity were tested on reference and environmental strains. Ruggedness, lot-to-lot consistency, and stability of the reagents kit were also studied.

RESULTS

Enumeration of L. pneumophila by MICA Legionella was statistically equivalent to ISO 11731:2017 in both matrixes. In cooling tower waters, MICA Legionella showed better sensitivity than ISO 11731:2017. It presented a 94% sensitivity and a 97% specificity.

CONCLUSION

MICA Legionella is a highly sensitive and specific method for culturable L. pneumophila enumeration. It presents, in 48 hours, equivalent or better results than ISO 11731:2017. Its protocol is robust to variations. Its reagents kit is stable for up to 18 months.

HIGHLIGHTS: MICA Legionella is a robust and reliable method for the enumeration of culturable L. pneumophila in domestic and cooling tower water. It reduces significantly the number of sample pretreatments required in ISO 11731:2017. Automatic identification and enumeration of L. pneumophila microcolonies eliminates the requirement to have skilled analysts and limits the results variability. It also greatly reduces the time to results to 48 h instead of 7-10 days with ISO 11731:2017 while providing statistically equivalent results.

Для цитирования: Passot Fanny, Peslier Sabine, Benzinger M Joseph, Blackburn Jonathan, Thompson Wesley, Bastin Benjamin, Dumont Audrey, Dukan Sam. Validation of MICA Legionella for Enumeration of Legionella pneumophila in Sanitary Waters and Cooling Tower Waters: AOAC Performance Tested MethodSM 032201.. Journal of AOAC International. 2023. DOI: 10.1093/jaoacint/qsac150.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.