Development and in-vitro characterization of a novel fetal vesicoamniotic shunt - The Vortex shunt.
Аннотация
OBJECTIVES
To develop and test a novel vesicoamniotic shunt (VAS) to treat fetal lower urinary tract obstruction (LUTO), decrease dislodgement and optimize shunt deployment in-vitro.
METHODS
Vesicoamniotic shunt design objectives included: (1) robust and atraumatic fixation elements, (2) kink resistant conduit to adjust to fetal movement and growth, (3) one-way pressure valve to facilitate bladder cycling, and (4) echogenic deployment visualization aids. The force to dislodge the novel Vortex shunt was compared with existing commercially available shunts in a bench-top porcine bladder model. Sonographic echogenicity was evaluated with ultrasound-guided deployment, and the shunt valve pressure measured.
RESULTS
A prototype novel Vortex shunt was developed using braided nitinol "umbrella-type" ends with a kink-resistant stem incorporating an internal one-way valve. The peak force required to dislodge the Vortex shunt was significantly higher than commercially available shunts (p < 0.01). Shunt deployment in the bench-top model was easily confirmed with ultrasound guidance and the brisk decompression of the inflated porcine bladder thereafter. In-vitro valve gauge pressure testing mirrored bladder pressures in human LUTO cases.
CONCLUSION
In-vitro testing shows that the Vortex shunt may improve deployment, sonographic visualization, kink resistance, and dynamic size adjustment. Validation in preclinical animal models are warranted and currently underway.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.