Аннотация

OBJECTIVE

To validate a novel device termed the equinometer, which is designed to accurately measure ankle dorsiflexion.

DESIGN

Test retest reliability analysis using serial measurements of dorsiflexion endpoint in a group of normal individuals.

SETTING

Motion analysis laboratory.

PARTICIPANTS

Ten healthy individuals.

PRIMARY OUTCOME MEASURE: The mean and standard deviation of the absolute difference in dorsiflexion endpoint for the group.

RESULTS

The mean absolute change in dorsiflexion endpoint for the group was 0.45 degrees with a standard deviation of 0.43 degrees.

CONCLUSIONS

With the use of the device described, the mean change in dorsiflexion endpoint was well within acceptable clinical limits. The reliability of measurements obtained with the equinometer exceeds that which has been published with other techniques and devices.

Для цитирования: Weaver K, Price R, Czerniecki J, Sangeorzan B. Design and validation of an instrument package designed to increase the reliability of ankle range of motion measurements.. Journal of rehabilitation research and development. 2001.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.