Artificial Intelligence Iterative Reconstruction for Dose Reduction in Pediatric Chest CT: A Clinical Assessment via Below 3 Years Patients With Congenital Heart Disease.
Аннотация
PURPOSE: To assess the performance of a newly introduced deep learning-based reconstruction algorithm, namely the artificial intelligence iterative reconstruction (AIIR), in reducing the dose of pediatric chest CT by using the image data of below 3-year-old patients with congenital heart disease (CHD).
MATERIALS AND
METHODS
The lung image available from routine-dose cardiac CT angiography (CTA) on below 3 years patients with CHD was employed as a reference for evaluating the paired low-dose chest CT. A total of 191 subjects were prospectively enrolled, where the dose for chest CT was reduced to ~0.1 mSv while the cardiac CTA protocol was kept unchanged. The low-dose chest CT images, obtained with the AIIR and the hybrid iterative reconstruction (HIR), were compared in image quality, ie, overall image quality and lung structure depiction, and in diagnostic performance, ie, severity assessment of pneumonia and airway stenosis.
RESULTS
Compared with the reference, lung image quality was not found significantly different on low-dose AIIR images (all P >0.05) but obviously inferior with the HIR (all P <0.05). Compared with the HIR, low-dose AIIR images also achieved a closer pneumonia severity index (AIIR 4.32±3.82 vs. Ref 4.37±3.84, P >0.05; HIR 5.12±4.06 vs. Ref 4.37±3.84, P <0.05) and airway stenosis grading (consistently graded: AIIR 88.5% vs. HIR 56.5% ) to the reference.
CONCLUSIONS
AIIR has the potential for large dose reduction in chest CT of patients below 3 years of age while preserving image quality and achieving diagnostic results nearly equivalent to routine dose scans.