MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1177/15385744251332765

Медицинская библиотека · Vascular and endovascular surgery · 2025

Shattered by Light: Catheter Fractures and the Hidden Danger of Angio-Seal Device Failure in the Femoral Artery.

Kotaki Rouzbeh, Shastri Ravi, Ghasemi-Rad Mohammad

Клинический случайПисьмо в редакциюПеревод в очереди

Аннотация

We report a case of a 77-year-old male with pancreatic cancer and thrombocytopenia who presented with acute stroke symptoms and underwent successful endovascular thrombectomy. During femoral artery closure, an 8-F Angio-Seal device fractured, leaving catheter fragments in the right femoral artery, confirmed on imaging. Device analysis revealed oxidation-induced brittleness, leading to sheath fragmentation. This case highlights a rare complication of Angio-Seal devices, suggesting potential material vulnerabilities that warrant further investigation.

Для цитирования: Kotaki Rouzbeh, Shastri Ravi, Ghasemi-Rad Mohammad. Shattered by Light: Catheter Fractures and the Hidden Danger of Angio-Seal Device Failure in the Femoral Artery.. Vascular and endovascular surgery. 2025. DOI: 10.1177/15385744251332765.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.