Portable dynamic ultrasonography is a useful tool for the evaluation of suspected syndesmotic instability: a cadaveric study.

Hagemeijer N C, Lubberts B, Saengsin J, Bhimani R, Sato G, Waryasz G R, Kerkhoffs G M M J, DiGiovanni C W, Guss D

СтатьяПеревод в очереди

Аннотация

PURPOSE: Portable ultrasonography (P-US) is increasingly used to diagnose syndesmotic instability. The aim of this study was to evaluate syndesmotic instability by measuring the distal tibiofibular clear space (TFCS) in a cadaveric model using P-US with progressive stages of syndesmotic ligamentous transection under external rotation stress.

METHODS

Ten fresh lower leg cadaveric specimens amputated above the proximal tibiofibular joint were used. Using P-US, the TFCS was evaluated in the intact stage and after progressive sectioning of the (1) anterior-inferior tibiofibular ligament (AITFL), (2) interosseous ligament (IOL), and (3) posterior-inferior tibiofibular ligament (PITFL). The TFCS was measured in both the unstressed (0 Nm) state and with 4.5, 6.0, 7.5, and 9.0 Nm of external rotation stress using a bone hook placed on the first metatarsal bone at each stage of ligamentous transection stage using both P-US and fluoroscopy.

RESULTS

When assessed with P-US, partial syndesmotic injury encompassing the AITFL and IOL resulted in significant TFCS widening at 4.5 Nm of external rotation torque when compared to intact state with a TFCS-opening of 2.6 ± 2 mm, p = 0.01. In contrast, no significant differences in TFCS were detected using fluoroscopy. Only a moderate correlation was found between P-US and fluoroscopy.

CONCLUSION

P-US is a useful tool in diagnosing syndesmotic instability during external rotation stress examination. TFCS-opening increased as additional ligaments of the syndesmosis were transected, and application of 4.5 Nm torque was sufficient to detect a difference of 2.6 mm after the IOL cut.

Для цитирования: Hagemeijer N C, Lubberts B, Saengsin J, Bhimani R, Sato G, Waryasz G R, Kerkhoffs G M M J, DiGiovanni C W, Guss D. Portable dynamic ultrasonography is a useful tool for the evaluation of suspected syndesmotic instability: a cadaveric study.. Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA. 2022. DOI: 10.1007/s00167-022-07058-4.