Increasing efficiency in the ACR Digital Mammography QC Manual's DBT Z-resolution evaluation.

Fisher Ryan F, Bevins Nicholas B, Hulme Katie W, Schaeffer Colin

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The Z-resolution evaluation in the ACR Digital Mammography (DM) QC manual measures a pseudo full width half max (pFWHM) from speck group signals through slices of the phantom for comparison to a baseline value.

PURPOSE: The evaluation requires measurement of seven ROIs on up to nine slices, for a potential of up to 63 ROIs. This work evaluates whether measuring fewer ROIs would produce similar results.

METHODS

Z-resolution data from 468 surveys from 147 mammography units were retrospectively analyzed. The pFWHM was calculated using the 6-speck mean (pFWHM6) as described in the ACR manual, as well as using each single speck (pFWHM1), and all combinations of a 2-speck mean (pFWHM2) using the center and each peripheral speck. Deviation from the respective initial measured pFWHMinitial was determined for all units with at least two surveys.

RESULTS

The pFWHM1 had an average absolute error of 6.0% relative to pFWHM6 across all surveys, compared to 3.9% for pFWHM2. For the 111 units with longitudinal data, the mean absolute deviation from pFWHMinitial was 7.8% for pFWHM1, compared to 5.6% for pFWHM2, and 5.1% for pFWHM6. For longitudinal testing, pFWHM6 produced two failures that were related to major component replacement. Both pFWHM1 and pFWHM2 methods caught those failures for a false negative rate of 0%. However, pFWHM1 had a false positive rate of 12.9%, while pFWHM2's was only 2.8%. No positive test from any method was attributable to actual equipment failure.

CONCLUSION

A single-speck pFWHM for ACR DM QC resulted in high variability and frequent unit failures for the Z-resolution evaluation. However, a 2-speck pFWHM produced results clinically equivalent to the 6-speck method. Using the 2-speck method as an alternative methodology for Z-resolution evaluation achieves similar results with 57% fewer measurements, increasing the efficiency of testing.

Для цитирования: Fisher Ryan F, Bevins Nicholas B, Hulme Katie W, Schaeffer Colin. Increasing efficiency in the ACR Digital Mammography QC Manual's DBT Z-resolution evaluation.. Journal of applied clinical medical physics. 2026. DOI: 10.1002/acm2.70619.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.