MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1186/s12882-023-03260-9

Медицинская библиотека · BMC nephrology · 2023

Rapid implementation of an emergency on-site CKRT dialysate production system during the COVID-19 pandemic.

Teixeira J Pedro, Saa Lisa, Kaucher Kevin A, Villanueva Ruben D, Shieh Michelle, Baca Crystal R, Harmon Brittany, Owen Zanna J, Mendez Majalca Ismael, Schmidt Darren W, Singh Namita, Shaffi Saeed K, Xu Zhi Q, Roha Thomas, Mitchell Jessica A, Demirjian Sevag, Argyropoulos Christos P

СтатьяПеревод в очереди

Аннотация

BACKGROUND

On December 29, 2021, during the delta wave of the Coronavirus Disease 2019 (COVID-19) pandemic, the stock of premanufactured solutions used for continuous kidney replacement therapy (CKRT) at the University of New Mexico Hospital (UNMH) was nearly exhausted with no resupply anticipated due to supply chain disruptions. Within hours, a backup plan, devised and tested 18 months prior, to locally produce CKRT dialysate was implemented. This report describes the emergency implementation and outcomes of this on-site CKRT dialysate production system.

METHODS

This is a single-center retrospective case series and narrative report describing and reporting the outcomes of the implementation of an on-site CKRT dialysate production system. All adults treated with locally produced CKRT dialysate in December 2021 and January 2022 at UNMH were included. CKRT dialysate was produced locally using intermittent hemodialysis machines, hemodialysis concentrate, sterile parenteral nutrition bags, and connectors made of 3-D printed biocompatible rigid material. Outcomes analyzed included dialysate testing for composition and microbiologic contamination, CKRT prescription components, patient mortality, sequential organ failure assessment (SOFA) scores, and catheter-associated bloodstream infections (CLABSIs).

RESULTS

Over 13 days, 22 patients were treated with 3,645 L of locally produced dialysate with a mean dose of 20.0 mL/kg/h. Fluid sample testing at 48 h revealed appropriate electrolyte composition and endotoxin levels and bacterial colony counts at or below the lower limit of detection. No CLABSIs occurred within 7 days of exposure to locally produced dialysate. In-hospital mortality was 81.8% and 28-day mortality was 68.2%, though illness severity was high, with a mean SOFA score of 14.5.

CONCLUSIONS

Though producing CKRT fluid with IHD machines is not novel, this report represents the first description of the rapid and successful implementation of a backup plan for local CKRT dialysate production at a large academic medical center in the U.S. during the COVID-19 pandemic. Though conclusions are limited by the retrospective design and limited sample size of our analysis, our experience could serve as a guide for other centers navigating similar severe supply constraints in the future.

Для цитирования: Teixeira J Pedro, Saa Lisa, Kaucher Kevin A, Villanueva Ruben D, Shieh Michelle, Baca Crystal R, Harmon Brittany, Owen Zanna J, Mendez Majalca Ismael, Schmidt Darren W, Singh Namita, Shaffi Saeed K, Xu Zhi Q, Roha Thomas, Mitchell Jessica A, Demirjian Sevag, Argyropoulos Christos P. Rapid implementation of an emergency on-site CKRT dialysate production system during the COVID-19 pandemic.. BMC nephrology. 2023. DOI: 10.1186/s12882-023-03260-9.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.