Interference screw versus suture anchor fixation for subpectoral tenodesis of the proximal biceps tendon: a cadaveric study.

Golish S Raymond, Caldwell Paul E, Miller Mark D, Singanamala Naveen, Ranawat Anil S, Treme Gehron, Pearson Sara E, Costic Ryan, Sekiya Jon K

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

PURPOSE: The purpose of this study was to compare the biomechanical properties of 2 fixation methods for subpectoral proximal biceps tenodesis.

METHODS

In 9 matched pairs of cadaveric shoulders, an open subpectoral tenodesis was performed 1 cm proximal to the inferior border of the pectoralis major tendon by use of either an 8 x 12-mm Bio-Tenodesis screw (Arthrex, Naples, FL) with No. 2 FiberWire sutures (Arthrex) or a 5.5-mm Bio-Corkscrew double-loaded suture anchor (Arthrex) with No. 2 FiberWire sutures. The specimens were dissected and mounted in a material testing machine. Cyclic loading (20 to 60 N, 100 cycles, 0.5 mm/s, 5-N preload) was performed, followed by an unloaded 30-minute rest, a 5-N preload, and a load-to-failure protocol (1.25 mm/s) with a 100-lb load cell. Ultimate load (in Newtons), stiffness (in Newtons per millimeter), and modes of failure were recorded. Data were analyzed by use of paired t tests and Wilcoxon signed rank tests.

RESULTS

Proximal biceps tenodeses with Bio-Tenodesis screws had a significantly higher mean load to failure (169.6 +/- 50.5 N; range, 99.6 to 244.7 N) than those with Bio-Corkscrew suture anchors (68.5 +/- 33.0 N; range, 24.2 to 119.4 N) (P = .002). Bio-Tenodesis screws also had a significantly higher stiffness (34.1 +/- 9.0 N/mm; range, 20.6 to 48.9 N/mm) than Bio-Corkscrews (19.3 +/- 10.5; range, 5.9 to 32.9 N/mm) (P = .038).

CONCLUSIONS

In this cadaveric study the Bio-Tenodesis screw showed a statistically significantly higher load to failure and significantly higher stiffness than the Bio-Corkscrew anchor when used for tenodesis of the proximal biceps tendon in a subpectoral location.

CLINICAL RELEVANCE: Biomechanical comparison of these 2 fixation techniques provides information on stiffness and load to failure of alternate fixation methods.

Для цитирования: Golish S Raymond, Caldwell Paul E, Miller Mark D, Singanamala Naveen, Ranawat Anil S, Treme Gehron, Pearson Sara E, Costic Ryan, Sekiya Jon K. Interference screw versus suture anchor fixation for subpectoral tenodesis of the proximal biceps tendon: a cadaveric study.. Arthroscopy : the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association. 2008. DOI: 10.1016/j.arthro.2008.05.005.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.