DOI 10.1111/j.1445-2197.2010.05436.x

Медицинская библиотека · ANZ journal of surgery · 2010

Evaluating the safety of the Harmonic Scalpel around the recurrent laryngeal nerve.

Jiang Hang, Shen Hongliang, Jiang Daozhen, Zheng Xiangmin, Zhang Wei, Lu Lei, Jiang Zhiguo, Qiu Ming

Сравнительное исследованиеОценочное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

The Harmonic Scalpel (HS) is widely used in thyroidectomy. Determining the safety margin of using the HS near the recurrent laryngeal nerve (RLN) is helpful in preventing the injury to this nerve during thyroidectomy.

METHODS

The parameters of evoked electromyography (EEMG) of vocal muscles before and after using the HS at a power level of 3 were recorded in a rabbit model. Masson staining was used to determine lateral injury caused by incisions using the HS.

RESULTS

After the activated tip of the HS made contact with the RLN for ≥1 s or was placed 1 mm from the nerve for 3 or 5 s or 2 mm from the nerve for 5 s, significant changes were observed in the minimal stimulus current intensity threshold, the optimal stimulus current intensity threshold, the onset latency and the wave amplitude of EEMG. After the activated HS tip touched the RLN or was placed 1 mm to the nerve for ≥1 s or 2 mm to the nerve for 5 s, significant changes were found in peak latency. The thickness of injured lateral tissue was <1, 1-2 and >2 mm when using HS for 1, 3 and 5 s, respectively.

CONCLUSION

When used near the RLN at a power level of 3, the activated HS tip should be ≥2 mm from the nerve and the duration of incision should be ≤3 s.

Для цитирования: Jiang Hang, Shen Hongliang, Jiang Daozhen, Zheng Xiangmin, Zhang Wei, Lu Lei, Jiang Zhiguo, Qiu Ming. Evaluating the safety of the Harmonic Scalpel around the recurrent laryngeal nerve.. ANZ journal of surgery. 2010. DOI: 10.1111/j.1445-2197.2010.05436.x.