Using a Cast Iron Skillet to Resolve Percutaneous Ventricular Assist Device and Implantable Cardioverter-Defibrillator Electromagnetic Interference.
Клинический случайСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Shaheriyar Mohammed, Taimur Muhammad, Busby James C, Hussain Zuhayr, Taj Quratulaine Fatima, Hassan Sohail
Аннотация
BACKGROUND: Percutaneous ventricular assist device (pVAD)-induced electromagnetic interference (EMI) can disrupt implantable cardioverter-defibrillator (ICD) telemetry. We describe a novel mitigation strategy using a cast iron skillet.
CASE SUMMARY: An 81-year-old woman with an ICD underwent pVAD-assisted percutaneous coronary intervention for cardiogenic shock. Postprocedure interrogation for slow ventricular tachycardia failed due to significant EMI. Because hemodynamic instability precluded reducing pVAD flow to mitigate interference, a cast iron skillet was placed over the heart. This pseudo-Faraday cage successfully restored telemetry.
DISCUSSION: Although the "skillet method" aids durable left ventricular assist device interrogation, prior reports suggest inefficacy with pVADs, favoring flow reduction. We report the first successful use of a cast iron skillet to shield Impella EMI, offering a vital strategy for unstable patients unable to tolerate pump weaning.
TAKE-HOME MESSAGES: pVAD systems can generate EMI, disrupting ICD telemetry. When hemodynamic instability prevents lowering pVAD flow, a cast iron skillet can assist in shielding EMI and restoring communication.
CASE SUMMARY: An 81-year-old woman with an ICD underwent pVAD-assisted percutaneous coronary intervention for cardiogenic shock. Postprocedure interrogation for slow ventricular tachycardia failed due to significant EMI. Because hemodynamic instability precluded reducing pVAD flow to mitigate interference, a cast iron skillet was placed over the heart. This pseudo-Faraday cage successfully restored telemetry.
DISCUSSION: Although the "skillet method" aids durable left ventricular assist device interrogation, prior reports suggest inefficacy with pVADs, favoring flow reduction. We report the first successful use of a cast iron skillet to shield Impella EMI, offering a vital strategy for unstable patients unable to tolerate pump weaning.
TAKE-HOME MESSAGES: pVAD systems can generate EMI, disrupting ICD telemetry. When hemodynamic instability prevents lowering pVAD flow, a cast iron skillet can assist in shielding EMI and restoring communication.