DOI 10.1016/j.bjps.2018.05.050

Медицинская библиотека · Journal of plastic, reconstructive & aesthetic surgery : JPRAS · 2018

Assessment of MRI safety issues for stainless steel sutures used for microtia reconstruction.

Kraai T L, Loch R B, Shellock F G

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

Potential magnetic resonance imaging issues for stainless steel sutures used for microtia reconstruction could be clinically significant for safety and diagnostic yield considerations. Therefore, the purpose of this investigation was to assess magnetic resonance issues (magnetic field interactions, heating, and artifacts) for different types of stainless steel sutures used for microtia reconstruction.

METHODS

Small gauge, commonly used stainless steel sutures from four different manufacturers (5/0 Steelex, Aesculap/B, Braun Medical, Inc.; Nagata 38 Gauge Microtia Wire, Bear Medical Corporation; Auricular Reco Wire, Medicon Surgical Inc.; and 5-0 B&S 35 Surgical Steel Suture, Ethicon, Inc.) were tested using standardized ex vivo techniques to assess magnetic field interactions, heating, and artifacts at 3 Tesla. Before testing, the stainless steel sutures were configured in a manner same as that for cartilage reconstruction used to treat microtia.

RESULTS

Each stainless steel suture exhibited minor magnetic field interactions at 3 Tesla (translational attraction, deflection angle <10°, and no torque). Heating associated with a whole-body averaged specific absorption rate of 2.9 W/kg was not excessive (highest temperature changes, ≤1.8 °C). Artifacts were relatively minor in relation to the size and shape of each stainless steel suture (artifact size in relation to the size and shape of each stainless steel suture extending ≤5 mm).

CONCLUSIONS

The stainless steel sutures that underwent testing do not present additional risks to patients in a 3-Tesla or less magnetic field setting (i.e., magnetic resonance conditional). Artifacts for these sutures may only be an issue within close proximity to the reconstructed ear.

Для цитирования: Kraai T L, Loch R B, Shellock F G. Assessment of MRI safety issues for stainless steel sutures used for microtia reconstruction.. Journal of plastic, reconstructive & aesthetic surgery : JPRAS. 2018. DOI: 10.1016/j.bjps.2018.05.050.