Robotic Assisted Cleft Palate Repair Using Novel 3 mm Tools: A Reachability and Collision Analysis.

Maguire Glenn, Tang Ethan, Looi Thomas, Podolsky Dale

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

Cleft palate repair is technically challenging: visualization is reduced, delicate dissection is required, and instrument reachability is limited within the infant oral cavity using standard instruments. A 3 mm diameter pin-jointed wristed tool was developed that couples to a da Vinci Research Kit (dVRK) to perform infant cleft palate surgery.

METHODS

A reachability simulation analysis was performed to determine the number of collisions (tool-tool and tool-cavity) for tool tip positions within the oral cavity and the impact of dVRK trocar height. Tool reachability was measured using a reachability ratio of collision-free to total poses considered for a given oral cavity target point. Three physical experiments were performed using a high-fidelity cleft palate simulator: (1) the 3 mm tools were compared to existing 8 mm tools assessing visual field obstruction and the number of collisions during suturing tasks; (2) an end-to-end robotic assisted cleft palate repair was performed on the simulator using the new tool assessing collisions; and (3) the impact of trocar height on tool performance was investigated to validate the reachability simulation analysis.

RESULTS

The 3 mm tools demonstrated fifteen times fewer tool-cavity collisions and 21% less visual obstruction than the 8 mm tools. Average tool-cavity collisions per throw demonstrated an increasing trend posteriorly within the oral cavity (nasal mucosa: $R^{2}=0.344$, $p=0.013$; oral mucosa: $R^{2}=0.811$, $p< 0.001$); and a 31% increase ($R^{2}=0.988$, $p=0.069$) with increasing trocar height.

CONCLUSION/SIGNIFICANCE: Physical experiments validated tool functionality for cleft repair and the reachability analysis reveals increasing collision risk deeper within the oral cavity.

Для цитирования: Maguire Glenn, Tang Ethan, Looi Thomas, Podolsky Dale. Robotic Assisted Cleft Palate Repair Using Novel 3 mm Tools: A Reachability and Collision Analysis.. IEEE transactions on bio-medical engineering. 2025. DOI: 10.1109/tbme.2025.3534119.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.