DOI 10.1089/dia.2021.0521

Медицинская библиотека · Diabetes technology & therapeutics · 2022

Feasibility of Closed-Loop Insulin Delivery with a Pregnancy-Specific Zone Model Predictive Control Algorithm.

Ozaslan Basak, Levy Carol J, Kudva Yogish C, Pinsker Jordan E, O'Malley Grenye, Kaur Ravinder Jeet, Castorino Kristin, Levister Camilla, Trinidad Mari Charisse, Desjardins Donna, Church Mei Mei, Plesser Mitchell, McCrady-Spitzer Shelly, Ogyaadu Selassie, Nelson Kristen, Reid Corey, Deshpande Sunil, Kremers Walter K, Doyle Francis J, Rosenn Barak, Dassau Eyal

СтатьяМногоцентровое исследованиеПоддержка NIH: внешняя программаПеревод в очереди

Аннотация

Objective

Evaluating the feasibility of closed-loop insulin delivery with a zone model predictive control (zone-MPC) algorithm designed for pregnancy complicated by type 1 diabetes (T1D). Research Design and

Methods

Pregnant women with T1D from 14 to 32 weeks gestation already using continuous glucose monitor (CGM) augmented pump therapy were enrolled in a 2-day multicenter supervised outpatient study evaluating pregnancy-specific zone-MPC based closed-loop control (CLC) with the interoperable artificial pancreas system (iAPS) running on an unlocked smartphone. Meals and activities were unrestricted. The primary outcome was the CGM percentage of time between 63 and 140 mg/dL compared with participants' 1-week run-in period. Early (2-h) postprandial glucose control was also evaluated.

Results

Eleven participants completed the study (age: 30.6 ± 4.1 years; gestational age: 20.7 ± 3.5 weeks; weight: 76.5 ± 15.3 kg; hemoglobin A1c: 5.6% ± 0.5% at enrollment). No serious adverse events occurred. Compared with the 1-week run-in, there was an increased percentage of time in 63-140 mg/dL during supervised CLC (CLC: 81.5%, run-in: 64%, P = 0.007) with less time >140 mg/dL (CLC: 16.5%, run-in: 30.8%, P = 0.029) and time <63 mg/dL (CLC: 2.0%, run-in:5.2%, P = 0.039). There was also less time <54 mg/dL (CLC: 0.7%, run-in:1.6%, P = 0.030) and >180 mg/dL (CLC: 4.9%, run-in: 13.1%, P = 0.032). Overnight glucose control was comparable, except for less time >250 mg/dL (CLC: 0%, run-in:3.9%, P = 0.030) and lower glucose standard deviation (CLC: 23.8 mg/dL, run-in:42.8 mg/dL, P = 0.007) during CLC.

Conclusion

In this pilot study, use of the pregnancy-specific zone-MPC was feasible in pregnant women with T1D. Although the duration of our study was short and the number of participants was small, our findings add to the limited data available on the use of CLC systems during pregnancy (NCT04492566).

Для цитирования: Ozaslan Basak, Levy Carol J, Kudva Yogish C, Pinsker Jordan E, O'Malley Grenye, Kaur Ravinder Jeet, Castorino Kristin, Levister Camilla, Trinidad Mari Charisse, Desjardins Donna, Church Mei Mei, Plesser Mitchell, McCrady-Spitzer Shelly, Ogyaadu Selassie, Nelson Kristen, Reid Corey, Deshpande Sunil, Kremers Walter K, Doyle Francis J, Rosenn Barak, Dassau Eyal. Feasibility of Closed-Loop Insulin Delivery with a Pregnancy-Specific Zone Model Predictive Control Algorithm.. Diabetes technology & therapeutics. 2022. DOI: 10.1089/dia.2021.0521.