DOI 10.5090/jcs.21.008

Медицинская библиотека · Journal of chest surgery · 2021

Development of a Screw-Crane System for Pre-Lifting the Sternal Depression in Pectus Excavatum Repair: A Test of Mechanical Properties for the Feasibility of a New Concept.

Park Hyung Joo, Rim Gongmin

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Pre-lifting of the sternum marked a major turning point in pectus excavatum repair. The author developed the crane technique in 2002 and successfully applied it to more than 2,000 cases using sternal wire stitching. However, blind sternal suturing limited the use of the wire-stitch crane. We propose a novel screw for sternal lifting as a new tool for the crane technique.

METHODS

We developed a screw system strong enough to withstand the pressure needed for sternum lifting. The screw was designed to have a broader thread to hold the bony tissue securely. The screw's sustaining power was tested using the torsion, driving torque, and axial pull-out tests in a polyurethane block and ex-vivo porcine sternum.

RESULTS

The screws were easily driven into the sternum, and the head of the screw was connectable to the table-mounted retractor. In the torsion test, the 2° offset torsional yield was 4.53 N·m (reference value, 1 N·m). In the polyurethane block driving torque test, the maximum torque was 0.98 N·m (reference value, 0.70 N·m). The axial pull-out test was 446 N (reference value, 100 N). The maximum pull-out resistance in the ex-vivo porcine sternum model was 1,516 N.

CONCLUSION

The screw crane was strong enough to sustain the chest wall weight to be lifted. Thus, the screws could effectively replace the sternal wire stitching in crane pre-lifting of the sternum. We expect that application of the screw-crane will be easy and that it will improve the safety and success rate of pectus repair surgery.

Для цитирования: Park Hyung Joo, Rim Gongmin. Development of a Screw-Crane System for Pre-Lifting the Sternal Depression in Pectus Excavatum Repair: A Test of Mechanical Properties for the Feasibility of a New Concept.. Journal of chest surgery. 2021. DOI: 10.5090/jcs.21.008.