Lloyd Michael S, Reddy Vivek Y, Roberts Paul, Doshi Rahul N, Wright David L, Boersma Lucas V A, Friedman Paul A, Neuzil Petr, Blomström-Lundqvist Carina, Bongiorni Maria Grazia, Burke Martin C, Gras Daniel, Kutalek Steven P, Marijon Eloi, Tolosana Jose María, Amin Anish K, Epstein Laurence M, Aasbo Johan D, Callahan Thomas D, Brisben Amy J, West Julie, Matznick Elizabeth, Speakman Benjamin, Bachman Tara N, Mont Lluís, Knops Reinoud E

Клиническое исследованиеСтатьяМногоцентровое исследованиеПеревод в очереди

Аннотация

BACKGROUND

MODULAR antitachycardia pacing (ATP), a multicenter, international trial, assesses a modular cardiac rhythm management system: a subcutaneous implantable cardioverter defibrillator in wireless communication with a leadless pacemaker (LP) capable of pace-terminating ventricular tachycardia.

METHODS

Enrolees had one or more clinical risk factors for ventricular tachycardia and did not require chronic pacing. Complications included prespecified major LP system- and procedure-related complications, and any complication related to the LP, subcutaneous implantable cardioverter defibrillator, implantation, or study protocol. Survival analysis was performed to identify complication-free rates, therapy delivery, and all-cause mortality.

RESULTS

The 297 patients enrolled had an ejection fraction of 35±13%, 43% secondary prevention indications, and 59% with prior ventricular arrhythmias. Of 286 patients undergoing LP implantation (100% success), 251 patients completed 12-month follow-up. Mortality rate was 6%, with none related to the implant procedure. Median follow-up duration was 23.4 months (interquartile range, 17.9-28.1). The LP major complication-free rate was 97.2%, exceeding the performance goal. The overall LP+ subcutaneous implantable cardioverter defibrillator system-related complication-free rate was 88.5%. Appropriate tachyarrhythmia-therapy (ATP+shock) rates were 14.4%, and appropriate shock rates were 8.5%. Inappropriate total tachyarrhythmia therapy was 9.5% of which 8.5% were shocks. ATP was 67.3% successful in terminating ventricular arrhythmia episodes and accelerated ventricular arrhythmias in 10.1% of episodes. Overall therapy burden (ATP+shock) was 96/100 patient-years, of which 44/100 patient-years was for shock delivery.

CONCLUSIONS

One-year outcomes of the first modular pacing-defibrillator system reveal low system and LP complication rates and good ATP efficacy rates, suggesting that the modular cardiac rhythm management is a viable alternative to single-chamber implantable cardioverter defibrillators using low-energy pacing capability without the need for transvenous leads.

CLINICAL TRIAL REGISTRATION: URL: https://clinicaltrials.gov/; Unique identifier: NCT04798768.

Для цитирования: Lloyd Michael S, Reddy Vivek Y, Roberts Paul, Doshi Rahul N, Wright David L, Boersma Lucas V A, Friedman Paul A, Neuzil Petr, Blomström-Lundqvist Carina, Bongiorni Maria Grazia, Burke Martin C, Gras Daniel, Kutalek Steven P, Marijon Eloi, Tolosana Jose María, Amin Anish K, Epstein Laurence M, Aasbo Johan D, Callahan Thomas D, Brisben Amy J, West Julie, Matznick Elizabeth, Speakman Benjamin, Bachman Tara N, Mont Lluís, Knops Reinoud E. One-Year Outcomes of the MODULAR ATP Trial: A Novel Leadless Pacemaker in Wireless Communication With a Subcutaneous Implantable Cardioverter Defibrillator.. Circulation. Arrhythmia and electrophysiology. 2025. DOI: 10.1161/circep.125.014395.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.