An AV-synchronous leadless pacemaker with electrical far-field atrial sensing in a sheep model.
Аннотация
BACKGROUND
Ventricular leadless pacemakers offer a valuable alternative to conventional transvenous systems by markedly reducing complications related to leads and device pockets. However, current leadless ventricular devices either provide asynchronous pacing only or achieve atrioventricular synchrony (AVS) with variable consistency, largely influenced by the underlying sinus rate, the pacing indication, and the programmed pacing strategy.
OBJECTIVE
To evaluate the safety and efficacy of a new AVS leadless pacemaker system in a Good Laboratory Practice animal study.
METHODS
The LivIQ DX pacemaker, a sub-1-cc VDD implant with tine fixation and electrical far-field sensing of the P-wave, was implanted in 9 sheep via jugular access in a Good Laboratory Practice study. Safety was evaluated during the implant procedure and via regular blood panels, ambulatory pacemaker data collection, and necropsy examination at 12 weeks post-implantation. Efficacy was assessed using R-wave amplitude, pacing capture threshold, and pacing impedance. AVS was evaluated at rest and during a stress test.
RESULTS
The LivIQ was successfully implanted in all sheep with no serious adverse events. Electrical values were stable throughout the 12-week duration, with a mean P-wave amplitude of 71 ± 39 μV, R-wave amplitude of 9.9 ± 4.1 mV, pacing capture threshold of 0.8 ± 0.4 V at 0.25 ms, and pacing impedance of 869 ± 233 Ω. All subjects were 100% AV-synchronous at rest during atrial-sensed ventricular-paced tracking, and AVS was maintained up to 140 beats per minute during the stress test. There were no clinically significant observations at necropsy.
CONCLUSION
This study highlights the safe implantation and reliable electrical performance of a novel leadless pacemaker, demonstrating high AVS in a chronic sheep model.