DOI 10.3760/cma.j.issn.1001-9391.2017.10.020
[Simultaneous determination of trichloroethylene and trichloroethanol in blood by liquid-liquid extraction-gas chromatography].
Аннотация
Objective
To establish a method for determing the trichloroethylene(TCE)and trichloroethanol(TCOH)in blood samples by liquid-liquid extraction-gas chromatography with electron capture detector.
Methods
With this method,ether was used as extraction solvent and trichloromethane was used as an internal standard. The whole blood sample was extracted with ether, and dehydrated by anhydrous sodium sulfate. Then the analytes were separated on HP-5 capillary column(30m×0.32mm×0.15μm)and detected byECD.The retention time was for qualitative analysis and the internal standard was for quantitation.
Results
The standard curves of TCE and TCOH showed significant linearity between 95.5μg/L-7640.0μg/L(r=0.9997)and 19.0μg/L-1520.0μg/L(r=0.9992). The average recovery was 95.5%-103.6%.The intra-day and inter-day precisions(RSD)were 2.5%-6.8%(n=6)and 1.6%-4.3%(n=6) respectively. The detect limit of TCE and TCOH were 2.10 μg/L and 0.56μg/L(S/N=3)respectively.The blood can be kept 7 days at-20℃ refrigerator without significantly loss.
Conclusion
This method is proved to be simple,practical and highly sensitive. It can satisfy the request for the determination of blood samples of humans exposed to TCE.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.