Safety and Effectiveness of a Novel Tips Access Set with Steerable Cannula in a Swine Model.

Ding PengXu, Ma Yujia, Zhu Xiaoxia, Wu Yijie, Ong John, Liu Pu, Xiao Jiangqiang, Zhuge Yuzheng

СтатьяПеревод в очереди

Аннотация

PURPOSE: This study aimed to assess the safety, effectiveness, and feasibility of the Liverty™ transjugular intrahepatic portosystemic shunt (TIPS) access set, which has an ergonomic handle that allows for in situ cannula tip deflection and a distal steerable cannula angle, versus the COOK® Rosch-Uchida Transjugular Liver Access Set (RUPS-100) in healthy pigs.

METHODS

Twelve pigs randomly underwent TIPS with the Liverty™ set or the RUPS-100 set. Three interventionalists performed 4 TIPS procedures, 2 with each set. The primary outcome was procedural success, defined as successful establishment of the intrahepatic portosystemic shunt and stent placement.

RESULTS

The shunt was successfully established in 11 pigs. The procedural success was achieved in all 6 pigs in the Liverty™ group and 5 out of 6 pigs for the RUPS-100 group (Fisher exact test, P > 0.999). The mean duration of puncture was shorter in the Liverty™ group versus the RUPS-100 group (12.3 ± 4.5 min vs. 16.2 ± 8.5 min), but without significant statistical difference (two sample t test, P = 0.359). The cannula angle was adjusted 69% of passes in the Liverty™ group, which was significantly higher than that in the RUPS-100 group (12%, P = 0.004). Overall, the TIPS procedural performance was comparable between the groups. Both sets were safe. No intraabdominal hemorrhage, vascular injuries, tissue or organ injuries, porto-biliary fistula, biliary peritonitis, and infection or abscess occurred in either group.

CONCLUSION

The Liverty™ set is safe and has similar procedural metrics to the COOK® RUPS-100 set. It allows in situ adjustment of the angle of the stiffening cannula without increasing procedure time and lessens the occurrences of periprocedural complications.

Для цитирования: Ding PengXu, Ma Yujia, Zhu Xiaoxia, Wu Yijie, Ong John, Liu Pu, Xiao Jiangqiang, Zhuge Yuzheng. Safety and Effectiveness of a Novel Tips Access Set with Steerable Cannula in a Swine Model.. Cardiovascular and interventional radiology. 2023. DOI: 10.1007/s00270-023-03544-5.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.