Treatment of distal radius fractures with palmar plates and locking screws. Comparisson of two different types of plate.
Аннотация
Various plating systems are available to fix distal radius fractures, each with a specific design. The purpose of this study was to compare radiological outcome and complications of the Variable Angle LCP Plate 2.4-mm (DePuy Synthes) with the VariAx volar locking plate (Stryker). One hundred patients (103 wrists) operated on for a distal radius fracture were retrospectively reviewed with a mean follow-up of 3.5 years. Seventy-three wrists were treated with a DePuy Synthes plate and 30 with a VariAx plate. The overall complication rate was 32%. Nineteen cases underwent revision surgery, 18 had malunion and 3 complex regional pain syndrome. Complicaton rate was 43% with DePuy Synthes plates and 27% with Variax plates, but the difference was not significant.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.