Optimal Selection of the Effective Echo Time (TEeff ) for T2-Weighted Fast-Spin-Echo MRI of the Prostate at 3.0 T: Effect on Lesion Conspicuity.

Kim Dahan, Pabi Ronard, Borisch Eric A, Froemming Adam T, Grimm Roger C, Hassanzadeh Sara, Kawashima Akira, Kido Ayumu, Takahashi Naoki, Thomas John V, Riederer Stephen J

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Clinical T 2-weighted (T2W) MRI of prostate cancer (PCa) usually implements a 2D fast-spin-echo (FSE) sequence, but understanding the echo-time-dependent contrast behavior is not straightforward due to the complicated FSE signal evolution. Consequently, consensus-based recommendations on the optimal effective echo time (TEeff) to maximize contrast are lacking.

PURPOSE: To investigate the effect of TEeff on PCa lesion contrast in prostate T2W MRI.

STUDY TYPE: Phantom and in vivo prospective study.

SUBJECTS: A standard MRI phantom and 53 prostate cancer patients (29 for image evaluation and 24 for quantitative lesion analysis).

FIELD STRENGTH AND SEQUENCE: Multi-slice T2W FSE sequence at 3 T acquired at multiple TEeff values.

ASSESSMENT: Three readers (13, 25, and 27 years' experience) evaluated TEeff = 100 ms versus 150 ms on diagnostic quality and reader preference. The effective T 2 relaxation rates (T 2,eff) were measured using FSE sequences in the phantom and used to predict the optimal TEeff maximizing T2W contrasts. From multi-TEeff scans of subjects with suspected PCa, T 2,eff values of PCa lesions and healthy tissue were measured and used to calculate the optimal TEeff maximizing the lesion contrast. In vivo contrast metrics were measured as a function of TEeff.

STATISTICAL TESTS: Wilcoxon test of reader scores, with p < 0.05 defining statistical significance.

RESULTS

No statistically significant difference was found in diagnostic quality between TEeff = 150 ms and 100 ms images, but all three readers significantly favored TEeff = 150 ms. Phantom experiments demonstrated that TEeff-dependent FSE signal progression can be characterized by T 2,eff values typically 40% longer than actual T 2 values, allowing accurate prediction of optimal TEeff within 2%-5%. Applied to PCa imaging, in vivo measurements indicated that effective T 2 relaxation times of the prostate are on average 1.5× longer than published T 2 values.

DATA

CONCLUSION

The benefits of TEeff = 150 ms versus 100 ms in prostate T2W FSE were supported by reader preference, phantom experiments, in vivo measurements, and analytic optimizations.

EVIDENCE LEVEL: 2.

TECHNICAL EFFICACY: 2.

Для цитирования: Kim Dahan, Pabi Ronard, Borisch Eric A, Froemming Adam T, Grimm Roger C, Hassanzadeh Sara, Kawashima Akira, Kido Ayumu, Takahashi Naoki, Thomas John V, Riederer Stephen J. Optimal Selection of the Effective Echo Time (TEeff ) for T2-Weighted Fast-Spin-Echo MRI of the Prostate at 3.0 T: Effect on Lesion Conspicuity.. Journal of magnetic resonance imaging : JMRI. 2026. DOI: 10.1002/jmri.70351.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.