m+Medelevator.RU
← Поиск публикаций
European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society · 2017

Development of a new assessment tool for cervical myelopathy using hand-tracking sensor: Part 1: validity and reliability.

СтатьяВалидационное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Alagha M Abdulhadi, Alagha Mahmoud A, Dunstan Eleanor, Sperwer Olaf, Timmins Kate A, Boszczyk Bronek M

Аннотация

PURPOSE: To assess the reliability and validity of a hand motion sensor, Leap Motion Controller (LMC), in the 15-s hand grip-and-release test, as compared against human inspection of an external digital camera recording.

METHODS: Fifty healthy participants were asked to fully grip-and-release their dominant hand as rapidly as possible for two trials with a 10-min rest in-between, while wearing a non-metal wrist splint. Each test lasted for 15 s, and a digital camera was used to film the anterolateral side of the hand on the first test. Three assessors counted the frequency of grip-and-release (G-R) cycles independently and in a blinded fashion. The average mean of the three was compared with that measured by LMC using the Bland-Altman method. Test-retest reliability was examined by comparing the two 15-s tests.

RESULTS: The mean number of G-R cycles recorded was: 47.8 ± 6.4 (test 1, video observer); 47.7 ± 6.5 (test 1, LMC); and 50.2 ± 6.5 (test 2, LMC). Bland-Altman indicated good agreement, with a low bias (0.15 cycles) and narrow limits of agreement. The ICC showed high inter-rater agreement and the coefficient of repeatability for the number of cycles was ±5.393, with a mean bias of 3.63.

CONCLUSIONS: LMC appears to be valid and reliable in the 15-s grip-and-release test. This serves as a first step towards the development of an objective myelopathy assessment device and platform for the assessment of neuromotor hand function in general. Further assessment in a clinical setting and to gauge healthy benchmark values is warranted.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты