Pacemaker therapy in low-birth-weight infants.
Аннотация
Infants born with complete atrioventricular block (CAVB) and fetal bradycardia are frequently born with low birth weight. Three low-birth-weight CAVB infants underwent temporary pacemaker implantation, followed by permanent single-chamber pacemaker implantation at median body weights of 1.7 and 3.2 kg, respectively. All infants caught up with their growth curves and had >3 years of estimated residual battery life. This two-stage strategy was successful in facilitating permanent pacemaker implantation in low-birth-weight babies. Placement of single-chamber pacemaker on the apex of the left ventricle appears to be associated with longer battery lifespan.