Reproducibility of risk figures in 2nd-trimester maternal serum screening for down syndrome: comparison of 2 laboratories.

Benn Peter A, Makowski Gregory S, Egan James F X, Wright Dave

Сравнительное исследованиеОценочное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

Analytical error affects 2nd-trimester maternal serum screening for Down syndrome risk estimation. We analyzed the between-laboratory reproducibility of risk estimates from 2 laboratories.

METHODS

Laboratory 1 used Bayer ACS180 immunoassays for alpha-fetoprotein (AFP) and human chorionic gonadotropin (hCG), Diagnostic Systems Laboratories (DSL) RIA for unconjugated estriol (uE3), and DSL enzyme immunoassay for inhibin-A (INH-A). Laboratory 2 used Beckman immunoassays for AFP, hCG, and uE3, and DSL enzyme immunoassay for INH-A. Analyte medians were separately established for each laboratory. We used the same computational algorithm for all risk calculations, and we used Monte Carlo methods for computer modeling.

RESULTS

For 462 samples tested, risk figures from the 2 laboratories differed >2-fold for 44.7%, >5-fold for 7.1%, and >10-fold for 1.7%. Between-laboratory differences in analytes were greatest for uE3 and INH-A. The screen-positive rates were 9.3% for laboratory 1 and 11.5% for laboratory 2, with a significant difference in the patients identified as screen-positive vs screen-negative (McNemar test, P<0.001). Computer modeling confirmed the large between-laboratory risk differences.

CONCLUSION

Differences in performance of assays and laboratory procedures can have a large effect on patient-specific risks. Screening laboratories should minimize test imprecision and ensure that each assay performs in a manner similar to that assumed in the risk computational algorithm.

Для цитирования: Benn Peter A, Makowski Gregory S, Egan James F X, Wright Dave. Reproducibility of risk figures in 2nd-trimester maternal serum screening for down syndrome: comparison of 2 laboratories.. Clinical chemistry. 2006. DOI: 10.1373/clinchem.2006.068783.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.