MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s00423-009-0503-y

Медицинская библиотека · Langenbeck's archives of surgery · 2009

A new anchor electrode design for continuous neuromonitoring of the recurrent laryngeal nerve by vagal nerve stimulations.

Schneider Rick, Przybyl Joanna, Hermann Michael, Hauss Johann, Jonas Sven, Leinung Steffen

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

PURPOSE: Intraoperative neuromonitoring has the limitation that the recurrent laryngeal nerve (RLN) is still at risk for damage between two stimulations with a handheld bipolar stimulation electrode. The purpose of this study was to establish the vagal anchor electrode for real-time monitoring of the RLN in surgical routine and to be alerted to imminent nerve failure by electromyography (EMG) signal analysis whereby the nerve damage becomes reversible.

PATIENTS AND

METHODS

This fully implantable electrode has been used in addition to a conventional handheld bipolar stimulation electrode during thyroid surgery on 45 consecutive patients (78 nerves at risk) stratified to low- and high-risk groups. The signal analysis was performed as real-time audio/video feedback by the use of a new multichannel EMG system.

RESULTS

No complications were attributable to the use of the anchor electrode. The mean delay to place the anchor electrode was 1.45 min, whereas the mean stimulation time of the vagus nerve was 38 min. Stable and repeatable signals were evocable in all cases with one exception. No permanent RLN paralyses occurred in this study.

CONCLUSIONS

The vagal anchor electrode is safe and easy to use. It allows continuous neuromonitoring without any threats. The new technique will provide more security, especially during preparation steps on the RLN that are difficult for the surgeon.

Для цитирования: Schneider Rick, Przybyl Joanna, Hermann Michael, Hauss Johann, Jonas Sven, Leinung Steffen. A new anchor electrode design for continuous neuromonitoring of the recurrent laryngeal nerve by vagal nerve stimulations.. Langenbeck's archives of surgery. 2009. DOI: 10.1007/s00423-009-0503-y.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.