MedElevator Войти в исследования
← Поиск публикаций
Медицинская библиотека · Undersea & hyperbaric medicine : journal of the Undersea and Hyperbaric Medical Society, Inc · 2026

A Non-Invasive Gas Exchange Monitor To Assess Swimming-Induced Pulmonary Edema.

Ahmed Yousef, Lindholm Peter, Bartlett Nick, Tripp Michael, Boswell Gil, Volk Charles

СтатьяНаблюдательное исследованиеПеревод в очереди

Аннотация

INTRODUCTION

Swimming-induced pulmonary edema (SIPE) occurs in physically demanding environments. SIPE has been identified among U.S. Naval Special Warfare (NSW) candidates, risking trainees' lives or, less severely, preventing them from completing training. Diagnosis is based on clinical examination and chest X-ray. The MediPines gas exchange monitor AGM-100 is a noninvasive diagnostic tool approved by the FDA and used in hospital settings. This study aimed to assess noninvasive physiological parameters and evaluate the use of the AGM-100 in NSW candidates with SIPE.

METHODS

In this observational study, eighteen subjects were tested (nine with confirmed SIPE). Data on a noninvasive surrogate for oxygen deficit, based on end-tidal oxygen (EtO₂),carbon dioxide (EtCO₂), and SpO₂ were collected. Chest X-rays (CXR) were graded for interstitial and airspace edema.

RESULTS

Subjects with confirmed SIPE displayed an oxygen deficit (28.1 ± 12.9 mm Hg), while eight of nine subjects without SIPE did not (3.1 ± 8.79 mm Hg, p < 0.001). There were no differences in EtCO₂. SpO₂ was lower in individuals with confirmed SIPE (95.1 ± 2.23%) compared to controls/unconfirmed SIPE (98.1 ± 0.31%, p-value < 0.0001). The group with the oxygen deficit had a correspondingly worse edema grade on CXR.

DISCUSSION: These limited results showed a possible correlation between O₂ deficit and SIPE, while the AGM-100 performed reliably in this prehospital setting. Future higher-powered studies are needed to substantiate these findings and the AGM-100's utility, possibly extending to other environments, such as combat field hospitals, for rapid diagnosis and severity assessment of pulmonary edema.

Для цитирования: Ahmed Yousef, Lindholm Peter, Bartlett Nick, Tripp Michael, Boswell Gil, Volk Charles. A Non-Invasive Gas Exchange Monitor To Assess Swimming-Induced Pulmonary Edema.. Undersea & hyperbaric medicine : journal of the Undersea and Hyperbaric Medical Society, Inc. 2026.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.