MedElevator Войти в исследования
← Поиск публикаций

DOI 10.3390/jcm15186948

Медицинская библиотека · Journal of clinical medicine · 2026

Robotic Arm Alignment System for Electrode Implantation in Invasive Epilepsy Monitoring: Clinical Experience and Accuracy in a Consecutive Patient Series with Comparison to the Literature.

Shawarba Julia, Zielke Johannes, Pataraia Ekaterina, Mayer Florian A, Roessler Karl

СтатьяПеревод в очереди

Аннотация

Objectives

Robot-assisted stereoelectroencephalography (SEEG) is increasingly used for invasive presurgical evaluation in drug-resistant epilepsy. Feasibility and accuracy data for robotic alignment systems remain limited. We analyzed our initial institutional experience and compared the results with the contemporary literature.

Methods

We retrospectively analyzed nine consecutive patients who underwent SEEG depth electrode implantation using the Brainlab Cirq robotic alignment system and intraoperative MRI for registration. Demographic and operative variables were extracted from institutional records. Implantation accuracy was measured on fused postoperative imaging as Euclidean entry point error (EPE) and target point error (TPE). A PubMed-focused literature screen was performed to identify studies reporting robotic SEEG implantation accuracy and/or efficiency metrics for comparison.

Results

Nine patients underwent implantation of 69 SEEG electrodes. Mean age was 21.7 ± 11.4 years. A mean of 7.7 ± 2.1 electrodes was implanted per patient. Mean operative time was 120.2 ± 31.9 min, corresponding to 16.0 ± 3.4 min per electrode. Mean EPE was 1.36 ± 0.60 mm and mean TPE was 1.79 ± 0.69 mm. Moreover, the maximum projected skull-entry angle was not independently associated with either EPE or TPE. In comparison with published robotic SEEG series and meta-analyses, our accuracy results fall within the established high-precision range and compare favorably with the most recent robotic-arm-specific data.

Conclusions

In this initial consecutive series, our robotic arm alignment system enabled accurate SEEG implantation with sub-2-mm mean target error. These findings support robotic arm systems as effective robotic alignment platforms for invasive epilepsy monitoring, while larger multicenter studies remain necessary to define comparative accuracy, learning curve effects, and complication profiles.

Для цитирования: Shawarba Julia, Zielke Johannes, Pataraia Ekaterina, Mayer Florian A, Roessler Karl. Robotic Arm Alignment System for Electrode Implantation in Invasive Epilepsy Monitoring: Clinical Experience and Accuracy in a Consecutive Patient Series with Comparison to the Literature.. Journal of clinical medicine. 2026. DOI: 10.3390/jcm15186948.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.