DOI 10.1093/ehjcr/ytae106

Медицинская библиотека · European heart journal. Case reports · 2024

Inadvertent septal perforation during conduction system pacing device implant: a case report.

Shtembari Jurgen, Shrestha Dhan Bahadur, Tung Roderick, Upadhyay Gaurav A

Клинический случайСтатьяПеревод в очереди

Аннотация

BACKGROUND

There has been recent growing interest in the use of conduction system pacing (CSP) for both bradycardia and heart failure indications. There remains a paucity of data, however, regarding complications related to the intraventricular septum associated with CSP implant and the management of these events.

CASE SUMMARY: We present a case of a patient with non-ischemic dilated cardiomyopathy presenting for cardiac resynchronization therapy in whom left bundle branch area pacing was complicated with interventricular septal perforation and managed intra-procedurally with repositioning of the lead to provide His bundle pacing (HBP) for QRS correction of underlying left bundle branch block. Post-procedure echocardiography did not show persistent ventricular septal defect. Left ventricular ejection fraction improved from 13% four months before implant to 30% at 32 months post-implant. Corrective HBP pacing thresholds showed a rise at 3-year follow-up.

DISCUSSION: Interventricular septal perforation during CSP is a possible complication during lead fixation. Pre-operative septal assessment with imaging can be helpful to provide important septal anatomical features. Septal perforation can be managed appropriately with lead repositioning intra-procedurally and close follow-up.

Для цитирования: Shtembari Jurgen, Shrestha Dhan Bahadur, Tung Roderick, Upadhyay Gaurav A. Inadvertent septal perforation during conduction system pacing device implant: a case report.. European heart journal. Case reports. 2024. DOI: 10.1093/ehjcr/ytae106.