Pulse oximetry signal loss during hypoxic episodes in preterm infants receiving automated oxygen control.

Langanky Lukas O, Kreutzer Karen B, Poets Christian F, Franz Axel R, Schwarz Christoph E

СтатьяРандомизированное исследованиеПеревод в очереди

Аннотация

UNLABELLED: The aim of this study was to analyze signal loss (SL) resulting from low signal quality of pulse oximetry-derived hemoglobin oxygen saturation (SpO2) measurements during prolonged hypoxemic episodes (pHE) in very preterm infants receiving automatic oxygen control (AOC). We did a post hoc analysis of a randomized crossover study of AOC, programmed to set FiO2 to "back-up FiO2" during SL. In 24 preterm infants (median (interquartile range)) gestational age 25.3 (24.6 to 25.6) weeks, recording time 12.7 h (12.2 to 13.6 h) per infant, we identified 76 pHEs (median duration 119 s (86 to 180 s)). In 50 (66%) pHEs, SL occurred for a median duration of 51 s (33 to 85 s) and at a median frequency of 2 (1 to 2) SL-periods per pHE. SpO2 before and after SL was similar (82% (76 to 88%) vs 82% (76 to 87%), p = 0.3)).

Conclusion

SL is common during pHE and must hence be considered in AOC-algorithm designs. Administering a "backup FiO2" (which reflects FiO2-requirements during normoxemia) during SL may prolong pHE with SL.  Trial registration: The study was registered at www.

CLINICALTRIALS: gov under the registration no. NCT03785899.

WHAT IS KNOWN: • Previous studies examined SpO2 signal loss (SL) during routine manual oxygen control being rare, but pronounced in lower SpO2 states. • Oxygen titration during SL is unlikely to be beneficial as SpO2 may recover to a normoxic range.

WHAT IS NEW: • Periods of low signal quality of SpO2 are common during pHEs and while supported with automated oxygen control (SPOC), FiO2 is set to a back-up value reflecting FiO2 requirements during normoxemia in response to SL, although SpO2 remained below target until signal recovery. • FiO2 overshoots following pHEs were rare during AOC and occurred with a delayed onset; therefore, increased FiO2 during SL does not necessarily lead to overshoots.

Для цитирования: Langanky Lukas O, Kreutzer Karen B, Poets Christian F, Franz Axel R, Schwarz Christoph E. Pulse oximetry signal loss during hypoxic episodes in preterm infants receiving automated oxygen control.. European journal of pediatrics. 2024. DOI: 10.1007/s00431-024-05549-9.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.