DOI 10.1086/647690

Медицинская библиотека · Infection control and hospital epidemiology · 1997

Bacterial killing ability of 10% ethylene oxide plus 90% hydrochlorofluorocarbon sterilizing gas.

Alfa M J, DeGagne P, Olson N

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVES

To use a serum and salt challenge in narrow-lumen carriers to evaluate a 10% ethylene oxide plus 90% hydrochlorofluorocarbon (EO-HCFC) sterilant mixture in a retrofitted 12/88 sterilizer as an alternative to the banned chlorofluorocarbon-ethylene oxide (EO) sterilant mixture.

DESIGN

An EO-HCFC sterilizing gas mixture in a retrofitted 12/88 sterilizer was compared to 100% ethylene oxide (100% EO) sterilizing gas to determine its relative ability to kill seven different bacteria (Escherichia coli, Pseudomonas aeruginosa, Enterococcus faecalis, Bacillus subtilis spores, Bacillus stearothermophilus spores, Bacillus circulans spores, and Mycobacterium chelonei) in the presence or absence of a combined 10% serum and 0.65% salt challenge using both penicylinders (PC) and long narrow-lumen (LU) carriers.

RESULTS

The EO-HCFC sterilant mixture (96% sterile carriers) was equivalent to the 100% EO sterilant (98% sterile carriers) for killing vegetative organisms, as well as spore suspensions, on the 27 PC and 27 LU carriers in the absence of serum and salt. In the presence of serum and salt, the EO-HCFC sterilant mixture was markedly better than the 100% EO sterilant at reducing the bacterial load on the 63 PC carriers (95% vs 62% sterile PC carriers, respectively), whereas both sterilizers were equivalent for the 63 LU carriers (49% vs 40% sterile LU carriers, respectively). Of the seven test organisms, E faecalis, B subtilis, B stearothermophilus, and B circulans were the most difficult to kill for both PC and LU carriers when serum and salt were present.

CONCLUSIONS

The data presented in this report indicate that the EO-HCFC sterilant mixture is an effective alternative for gas sterilization. Indeed, the efficiency of bacterial killing for the EO-HCFC sterilant mixture was similar to that achieved by the 12/88 EO-CFC sterilant mixture.

Для цитирования: Alfa M J, DeGagne P, Olson N. Bacterial killing ability of 10% ethylene oxide plus 90% hydrochlorofluorocarbon sterilizing gas.. Infection control and hospital epidemiology. 1997. DOI: 10.1086/647690.