Quantification of meclizine in human plasma by high performance liquid chromatography-mass spectrometry.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Wang Zhijun, Qian Shuai, Zhang Qizhi, Chow Moses S S
Аннотация
PURPOSE: Meclizine is an antihistamine and has been widely used for prophylactic treatment of motion sickness. To facilitate its pharmacokinetic study in human subjects, a high performance liquid chromatography-mass spectrometric method employing positive electrospray ionization was developed for the determination of meclizine concentration in human plasma.
METHODS: Meclizine together with the internal standard (flunarizine) was extracted from 0.1 ml of human plasma by protein precipitation using acetonitrile. The chromatography was performed using a Zorbax SB-C18 column (150 × 2.1mm, 5 μm, Agilent) with the mobile phase consisting of acetonitrile and 0.2% formic acid containing 2mM amino acetate. Multiple reaction monitoring was used for quantification. The validation of the method including sensitivity, linearity, reproducibility and stability was examined.
RESULTS: The lower limit of quantification (LLOQ) of the developed assay method for meclizine was 0.5 ng/ml and the linear calibration curve was acquired with R² > 0.99 between 0.5 and 200ng/ml. The intra-day and inter-day variation of the current assay was evaluated with the coefficient of variations (CVs%) within 12.92% at LLOQ and 7.15% for other quality control samples, whereas the mean accuracy ranged from 99.2% to 102.7%. The samples were stable under the storage conditions at least for a month.
CONCLUSION: The present method provides a robust, fast and sensitive analytical tool for meclizine in human plasma and has been successfully applied to a clinical pharmacokinetic study in 20 subjects.
METHODS: Meclizine together with the internal standard (flunarizine) was extracted from 0.1 ml of human plasma by protein precipitation using acetonitrile. The chromatography was performed using a Zorbax SB-C18 column (150 × 2.1mm, 5 μm, Agilent) with the mobile phase consisting of acetonitrile and 0.2% formic acid containing 2mM amino acetate. Multiple reaction monitoring was used for quantification. The validation of the method including sensitivity, linearity, reproducibility and stability was examined.
RESULTS: The lower limit of quantification (LLOQ) of the developed assay method for meclizine was 0.5 ng/ml and the linear calibration curve was acquired with R² > 0.99 between 0.5 and 200ng/ml. The intra-day and inter-day variation of the current assay was evaluated with the coefficient of variations (CVs%) within 12.92% at LLOQ and 7.15% for other quality control samples, whereas the mean accuracy ranged from 99.2% to 102.7%. The samples were stable under the storage conditions at least for a month.
CONCLUSION: The present method provides a robust, fast and sensitive analytical tool for meclizine in human plasma and has been successfully applied to a clinical pharmacokinetic study in 20 subjects.