MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jchromb.2010.11.022

Медицинская библиотека · Journal of chromatography. B, Analytical technologies in the biomedical and life sciences · 2010

Quantification of meclizine in human plasma by high performance liquid chromatography-mass spectrometry.

Wang Zhijun, Qian Shuai, Zhang Qizhi, Chow Moses S S

СтатьяПеревод в очереди

Аннотация

PURPOSE: Meclizine is an antihistamine and has been widely used for prophylactic treatment of motion sickness. To facilitate its pharmacokinetic study in human subjects, a high performance liquid chromatography-mass spectrometric method employing positive electrospray ionization was developed for the determination of meclizine concentration in human plasma.

METHODS

Meclizine together with the internal standard (flunarizine) was extracted from 0.1 ml of human plasma by protein precipitation using acetonitrile. The chromatography was performed using a Zorbax SB-C18 column (150 × 2.1mm, 5 μm, Agilent) with the mobile phase consisting of acetonitrile and 0.2% formic acid containing 2mM amino acetate. Multiple reaction monitoring was used for quantification. The validation of the method including sensitivity, linearity, reproducibility and stability was examined.

RESULTS

The lower limit of quantification (LLOQ) of the developed assay method for meclizine was 0.5 ng/ml and the linear calibration curve was acquired with R² > 0.99 between 0.5 and 200ng/ml. The intra-day and inter-day variation of the current assay was evaluated with the coefficient of variations (CVs%) within 12.92% at LLOQ and 7.15% for other quality control samples, whereas the mean accuracy ranged from 99.2% to 102.7%. The samples were stable under the storage conditions at least for a month.

CONCLUSION

The present method provides a robust, fast and sensitive analytical tool for meclizine in human plasma and has been successfully applied to a clinical pharmacokinetic study in 20 subjects.

Для цитирования: Wang Zhijun, Qian Shuai, Zhang Qizhi, Chow Moses S S. Quantification of meclizine in human plasma by high performance liquid chromatography-mass spectrometry.. Journal of chromatography. B, Analytical technologies in the biomedical and life sciences. 2010. DOI: 10.1016/j.jchromb.2010.11.022.