Flow confirmation study for central venous port in oncologic outpatient undergoing chemotherapy: evaluation of suspected system-related mechanical complications.

Sofue Keitaro, Arai Yasuaki, Takeuchi Yoshito, Sugimura Kazuro

Контролируемое клиническое исследованиеСтатьяПеревод в очереди

Аннотация

PURPOSE: To evaluate the efficacy and outcome of a flow confirmation study (FCS) in oncologic outpatients undergoing chemotherapy suspected of a central venous port (CVP) system-related mechanical complication.

MATERIALS AND

METHODS

A total of 66 patients (27 men, 39 women; mean age, 60 years) received FCS for the following reasons: prolonged infusion time during chemotherapy (n=32), inability to inject saline fluid (n=15), lateral neck and/or back pain (n=6), subcutaneous extravasation of anticancer drug (n=5), arm swelling (n=4), and inability to puncture the port (n=4). FCS consisted of examining the position of CVP, potential secondary shifts or fractures, and integrity of the system using contrast material through the port.

RESULTS

Of the 66 patients, 43 had an abnormal finding uncovered by FCS. The most frequent abnormal findings was catheter kinking (n=22). Explantation and reimplantation of the CVP system was required in 21 of the 66 patients. Remaining 45 patients were able continue using the CVP system after the FCS without any system malfunction.

CONCLUSION

FCS was effective for evaluating CVP system-related mechanical complications and was useful for deciding whether CVP system explantation and reimplantation was required.

Для цитирования: Sofue Keitaro, Arai Yasuaki, Takeuchi Yoshito, Sugimura Kazuro. Flow confirmation study for central venous port in oncologic outpatient undergoing chemotherapy: evaluation of suspected system-related mechanical complications.. European journal of radiology. 2013. DOI: 10.1016/j.ejrad.2013.07.003.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.