Huber Michaela, Loibl Markus, Eder Christoph, Kujat Richard, Nerlich Michael, Gehmert Sebastian
Аннотация
PURPOSE: This cadaver study investigated the temperature profile in the wrist joint and distal radioulnar joint (DRUJ) during radiofrequency energy (RFE) application for triangular fibrocartilage complex resection.
METHODS: An arthroscopic partial resection of the triangular fibrocartilage complex using monopolar and bipolar RFE was simulated in 14 cadaver limbs. The temperature was recorded simultaneously in the DRUJ and at 6 other anatomic locations of the wrist during RFE application.
RESULTS: The mean temperature in the DRUJ was 43.3 ± 8.2°C for the bipolar system in the ablation mode (60 W) and 30.4 ± 3.4°C for the monopolar system in the cut mode (20 W) after 30 seconds. The highest measured temperature in the DRUJ was 54.3°C for the bipolar system and 68.1°C for the monopolar system.
CONCLUSIONS: The application of RFE for debridement or resection of the triangular fibrocartilage complex in a clinical setting can induce peak temperatures that might cause damage to the cartilage of the DRUJ. Bipolar systems produce higher mean temperatures than monopolar devices.
CLINICAL RELEVANCE: RFE application increases the mean temperature in the DRUJ after 30 seconds to a level that may jeopardize cartilage tissue.
METHODS: An arthroscopic partial resection of the triangular fibrocartilage complex using monopolar and bipolar RFE was simulated in 14 cadaver limbs. The temperature was recorded simultaneously in the DRUJ and at 6 other anatomic locations of the wrist during RFE application.
RESULTS: The mean temperature in the DRUJ was 43.3 ± 8.2°C for the bipolar system in the ablation mode (60 W) and 30.4 ± 3.4°C for the monopolar system in the cut mode (20 W) after 30 seconds. The highest measured temperature in the DRUJ was 54.3°C for the bipolar system and 68.1°C for the monopolar system.
CONCLUSIONS: The application of RFE for debridement or resection of the triangular fibrocartilage complex in a clinical setting can induce peak temperatures that might cause damage to the cartilage of the DRUJ. Bipolar systems produce higher mean temperatures than monopolar devices.
CLINICAL RELEVANCE: RFE application increases the mean temperature in the DRUJ after 30 seconds to a level that may jeopardize cartilage tissue.