Atrial defibrillation thresholds of electrode configurations available to an atrioventricular defibrillator.

Benser M E, Walcott G P, Killingsworth C R, Girouard S D, Morris M M, Ideker R E

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

INTRODUCTION

Little investigation has been conducted to assess the atrial defibrillation thresholds of electrode configurations using electrodes designed for internal ventricular defibrillation (right ventricle [RV], superior vena cava [SVC], and pulse generator housing [Can]) combined with coronary sinus (CS) electrodes. We hypothesized that a CS-->SVC+Can electrode configuration would have a lower atrial defibrillation threshold than a standard configuration for defibrillation, RV-->SVC+Can. We also tested the atrial defibrillation thresholds of five other configurations.

METHODS AND

RESULTS

In 12 closed chest sheep, we situated a two-coil (RV, SVC) defibrillation catheter, a left-pectoral subcutaneous Can, and a CS lead. Atrial fibrillation was burst induced and maintained with continuous infusion of intrapericardial acetyl-beta-methylcholine chloride. Using fixed-tilt biphasic shocks, we determined the atrial defibrillation thresholds of seven test configurations in random order according to a multiple-reversal protocol. The peak voltage and delivered energy atrial defibrillation thresholds of CS-->SVC+Can (168+/-67 V, 2.68+/-2.40 J) were significantly lower than those of RV-->SVC+Can (215+/-88 V, 4.46+/-3.40 J). The atrial defibrillation thresholds of the other test configurations were RV+CS-->SVC+Can: 146+/-59 V, 1.92+/-1.45 J; RV-->CS+SVC+Can: 191+/-89 V, 3.53+/-3.19 J; CS-->SVC: 188+/-98 V, 3.77+/-4.14 J; SVC-->CS+ Can: 265+/-145 V, 7.37+/-9.12 J; and SVC-->Can: 516+/-209 V, 24.5+/-15.0 J.

CONCLUSIONS

The atrial defibrillation threshold of CS-->SVC+Can is significantly lower than that of RV-->SVC+Can. In addition, the low atrial defibrillation threshold of RV+CS-->SVC+Can merits further investigation. Based on corroboration of low atrial defibrillation thresholds of CS-based configurations in humans, physicians might consider using CS leads with atrioventricular defibrillators.

Для цитирования: Benser M E, Walcott G P, Killingsworth C R, Girouard S D, Morris M M, Ideker R E. Atrial defibrillation thresholds of electrode configurations available to an atrioventricular defibrillator.. Journal of cardiovascular electrophysiology. 2001. DOI: 10.1046/j.1540-8167.2001.00957.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.