Safety and Efficacy of Temperature-Guided Radiofrequency Ablation for Ventricular Tachycardia: A Multicenter Study.

Compagnucci Paolo, Huang Henry D, Bergonti Marco, Valeri Yari, Gasperetti Alessio, Schiavone Marco, Cespón-Fernández María, Hasegawa Kanae, Schillaci Vincenzo, Solimene Francesco, Cipolletta Laura, Parisi Quintino, Volpato Giovanni, Gellér László, Szegedi Nándor, Tondo Claudio, Sarkozy Andrea, Tandri Harikrishna, Casella Michela, Stevenson William G, Dello Russo Antonio

СтатьяМногоцентровое исследованиеНаблюдательное исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

A catheter platform allowing power- and temperature-controlled ablation (temperature-guided ablation) has recently been introduced. Clinical outcomes after ventricular tachycardia (VT) ablation using this technology remain uncertain.

OBJECTIVES

The goal of this study was to evaluate the safety and efficacy of temperature-guided VT ablation and compare the findings vs those from standard power-controlled ablation.

METHODS

This multicenter cohort study enrolled consecutive patients undergoing temperature-guided VT ablation using the QDOT MICRO catheter (Biosense Webster) at 8 referral centers in Europe and the United States (2021-2024). For comparison, a multicenter cohort of patients receiving power-controlled ablation with the THERMOCOOL SMARTTOUCH or SMARTTOUCH SurroundFlow (Biosense Webster) catheters was also included. The primary safety endpoint was procedure-related complications. The primary efficacy endpoint was sustained VT/ventricular fibrillation recurrence during long-term follow-up, assessed in the overall cohort and in propensity score-matched patients.

RESULTS

The study included 286 patients: 109 treated with temperature-guided ablation and 177 with power-controlled ablation. Ischemic cardiomyopathy was the predominant VT substrate (n = 125 [44%]); approximately one-half presented with electrical storm. Propensity score matching based on baseline clinical data yielded 101 pairs. Procedure-related complications were significantly lower with temperature-guided ablation (temperature-guided: n = 3 [3%]; power-controlled: n = 23 [13%]; P = 0.003), a result that remained significant after matching (temperature-guided: n = 3 [3%]; power-controlled: n = 12 [12%]; P = 0.028) and excluding vascular complications (temperature-guided: n = 1 [1%]; power-controlled: n = 14 [8%]; P = 0.012). Use of temperature-guided ablation was associated with lower odds of nonvascular complications in multivariable analysis (adjusted OR: 0.095; P = 0.028). Over a median follow-up of 24 months, VT/ventricular fibrillation recurrence rates were similar between groups in both overall (log-rank test, P = 0.570) and matched (log-rank test, P = 0.850) cohorts. Subgroup analyses showed no signal of heterogeneity.

CONCLUSIONS

In this multicenter VT ablation registry, a novel temperature-guided ablation modality showed favorable safety and efficacy relative to power-controlled ablation.

Для цитирования: Compagnucci Paolo, Huang Henry D, Bergonti Marco, Valeri Yari, Gasperetti Alessio, Schiavone Marco, Cespón-Fernández María, Hasegawa Kanae, Schillaci Vincenzo, Solimene Francesco, Cipolletta Laura, Parisi Quintino, Volpato Giovanni, Gellér László, Szegedi Nándor, Tondo Claudio, Sarkozy Andrea, Tandri Harikrishna, Casella Michela, Stevenson William G, Dello Russo Antonio. Safety and Efficacy of Temperature-Guided Radiofrequency Ablation for Ventricular Tachycardia: A Multicenter Study.. JACC. Clinical electrophysiology. 2026. DOI: 10.1016/j.jacep.2026.01.012.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.